Accuracy of mammography dosimetry in the era of the European Directive 2013/59/Euratom transposition.
Breast density
Calibration
Data accuracy
Mammography
Mean glandular dose
Radiation protection
Journal
European journal of radiology
ISSN: 1872-7727
Titre abrégé: Eur J Radiol
Pays: Ireland
ID NLM: 8106411
Informations de publication
Date de publication:
Jun 2020
Jun 2020
Historique:
received:
24
10
2019
revised:
30
01
2020
accepted:
28
03
2020
pubmed:
17
4
2020
medline:
8
1
2021
entrez:
17
4
2020
Statut:
ppublish
Résumé
To evaluate the impact of increasing levels of accuracy for mean glandular dose (MGD) evaluation in the era of the European Directive 2013/59/Euratom transposition. 4028 women who had a mammography examination by one of five mammography units using different detector technologies were included in this study. 16,006 images were processed by a software algorithm that determines breast glandularity quantitatively and uses this to estimate patient-specific MGD (psMGD). Entrance dose (ED) values and half value layers (HVLs) measured for each mammography system were collected to evaluate the effect of equipment calibration in psMGD calculation. The psMGD values adjusted for system calibration were compared with organ dose (OD) provided by manufacturers as image metadata. Overall median relative difference between calibrated psMGD and organ dose was below 3%, with larger differences for individual systems. The psMGD adjustment for system calibration was particularly useful for one system for which ED had an evident miscalibration issue. The mean difference between psMGD with calibration and organ dose provided by manufacturers was 4.1 %, ranging from -16.3 % to +24.5 %. The proportion of images for which organ dose was more than 10 % 'inaccurate' compared to psMGD was between 11 % and 46 %, depending on the mammography system. Patient-specific mean glandular dose, possibly adjusted for system calibration, allows more accurate individual breast dosimetry than what would be performed using organ dose provided by manufacturers. Conversely, definition of diagnostic reference levels could be achieved using either psMGD or organ dose.
Identifiants
pubmed: 32298958
pii: S0720-048X(20)30175-3
doi: 10.1016/j.ejrad.2020.108986
pii:
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
108986Commentaires et corrections
Type : CommentIn
Type : CommentIn
Informations de copyright
Copyright © 2020 Elsevier B.V. All rights reserved.