Physical characterization of a novel wireless DRX Plus 3543C using both a carbon nano tube (CNT) mobile x-ray system and a traditional x-ray system.
Journal
Physics in medicine and biology
ISSN: 1361-6560
Titre abrégé: Phys Med Biol
Pays: England
ID NLM: 0401220
Informations de publication
Date de publication:
11 06 2020
11 06 2020
Historique:
pubmed:
21
4
2020
medline:
12
9
2020
entrez:
21
4
2020
Statut:
epublish
Résumé
This work aims to characterize the novel DRX Plus 3543C detector in terms of detective quantum efficiency (DQE) using both a mobile x-ray system called Carestream DRX Revolution Nano and a traditional x-ray system (Carestream DRX Evolution). We used the commercial system DRX Revolution Nano, equipped with a new x-ray source based on CNT technology and field emission (FE) as the electron emitter (cathode). An innovative aspect of this device is its intrinsic selection of the focal spot size. We tested the system using three IEC-specified beam qualities (RQA3, 5 and 7) in terms of modulation transfer function (MTF), normalized noise power spectra (NNPS) and DQE as defined in the IEC 62220-1-1:2015. We compared the results obtained using DRX Revolution Nano and DRX Evolution with correlation and with Bland-Altman plots to study their agreement. RQA3 MTF is slightly lower than the RQA5 and 7 curves between 0.5 and 2.5 cycles mm