A tensed Tendyne.
Mitral
Tendyne
Transcatheter valve
implantation
valve
Journal
European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery
ISSN: 1873-734X
Titre abrégé: Eur J Cardiothorac Surg
Pays: Germany
ID NLM: 8804069
Informations de publication
Date de publication:
01 09 2020
01 09 2020
Historique:
received:
31
12
2019
revised:
04
03
2020
accepted:
09
03
2020
pubmed:
26
4
2020
medline:
22
6
2021
entrez:
26
4
2020
Statut:
ppublish
Résumé
We report a case of prosthesis dislodgement after transcatheter mitral valve replacement in an 85-year-old woman with chronic ischaemic heart failure. Two weeks after an initial successful implantation, she presented with a paravalvular leak associated with left ventricular outflow tract obstruction. Tether re-tensioning was performed and resolved the situation, but resulted in a deformation of the apical attachment zone into the left ventricle. Unfortunately, the patient finally expired from severe endocarditis. Proper anchoring is the main challenge for transcatheter mitral valve replacement techniques. Dislodgement of the prosthesis after transcatheter mitral valve replacement is an infrequent complication of the Tendyne® procedure. This case emphasizes the importance of assessing the quality of the myocardium at the implantation zone of the apical pad, and of prosthesis oversizing, especially if low-profile valves are chosen. .
Identifiants
pubmed: 32333016
pii: 5824970
doi: 10.1093/ejcts/ezaa109
doi:
Types de publication
Case Reports
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
651-653Informations de copyright
© The Author(s) 2020. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.