Résumé
The use of surface radiofrequency (RF) coils is common practice to boost sensitivity in (pre)clinical MRI. The number of transceive surface RF coils is rapidly growing due to the surge in cryogenically cooled RF technology and ultrahigh-field MRI. Consequently, there is an increasing need for effective correction of the excitation field ( We implemented the commonly used sensitivity correction and developed an empirical model-based method and a hybrid combination of both. Tests and validations were performed with a cryogenically cooled RF probe and a single-loop RF coil. Accuracy of SI quantification and T The three described correction methods achieved dramatic improvements in B This work demonstrates the efficacy of three B
Types de publication
Journal Article
Research Support, Non-U.S. Gov't
Langues
eng
Sous-ensembles de citation
IM
Pagination
2684-2701Informations de copyright
© 2020 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals LLC on behalf of International Society for Magnetic Resonance in Medicine.
Références
Constable R, Henkelman R. Contrast, resolution, and detectability in MR imaging. J Comput Assist Tomogr. 1991;15:297-303.
Fuderer M. The information content of MR images. IEEE Trans Med Imaging. 1988;7:368-380.
Redpath T. Signal-to-noise ratio in MRI. Br J Radiol. 1998;71:704-707.
Portnoy S, Kale S, Feintuch A, Tardif C, Pike G, Henkelman R. Information content of SNR/resolution trade-offs in three-dimensional magnetic resonance imaging. Med Phys. 2009;36:1442-1451.
Kale S, Chen X, Henkelman R. Trading off SNR and resolution in MR images. NMR Biomed. 2009;22:488-494.
Griswold MA, Jakob PM, Heidemann RM, et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002;47:1202-1210.
Pruessmann K, Weiger M, Scheidegger M, Boesiger P. SENSE: Sensitivity encoding for fast MRI. Magn Reson Med. 1999;42:952-962.
Lustig M, Donoho D, Pauly J. Sparse MRI: The application of compressed sensing for rapid MR imaging. Magn Reson Med. 2007;58:1182-1195.
Niendorf T, Barth M, Kober F, Trattnig S. From ultrahigh to extreme field magnetic resonance: Where physics, biology and medicine meet. Magn Reson Mater Phys, Biol Med. 2016;29:309-311.
Ladd M. The quest for higher sensitivity in MRI through higher magnetic fields. Zeitschrift für medizinische Physik. 2018;28:1-3.
Moser E, Laistler E, Schmitt F, Kontaxis G. Ultra-high field NMR and MRI-The role of magnet technology to Increase sensitivity and specificity. Front Phys. 2017;5:33. https://doi.org/10.3389/fphy.2017.0003333.
Axel L, Hayes C. Surface coil magnetic resonance imaging. Archives Internationales de Physiologie et de Biochimie. 1985;93:11-18.
Crowley M, Evelhoch J, Ackermann J. The surface-coil NMR receiver in the presence of homogeneous B1 excitation. J Magn Reson. 1985;64:20-31.
Hosseinnezhadian S, Frass-Kriegl R, Goluch-Roat S, et al. A flexible 12-channel transceiver array of transmission line resonators for 7 T MRI. J Magn Reson. 2018;296:47-59.
Pinkerton R, Barberi E, Menon R. Transceive surface coil array for magnetic resonance imaging of the human brain at 4 T. Magn Reson Med. 2005;54:499-503.
Pinkerton R, Near J, Barberi E, Menon R, Bartha R. Transceive surface coil array for MRI of the human prostate at 4T. Magn Reson Med. 2007;57:455-458.
Graessl A, Renz W, Hezel F, et al. Modular 32-channel transceiver coil array for cardiac MRI at 7.0T. Magn Reson Med. 2014;72:276-290.
Rietsch SHG, Pfaffenrot V, Bitz AK, et al. An 8-channel transceiver 7-channel receive RF coil setup for high SNR ultrahigh-field MRI of the shoulder at 7T. Med Phys. 2017;44:6195-6208.
He X, Ertürk MA, Grant A, et al. First in-vivo human imaging at 10.5T: Imaging the body at 447 MHz. Magn Reson Med. 2020;84:289-303. https://doi.org/10.1002/mrm.28131.
Ertürk M, Raaijmakers A, Adriany G, Ugurbil K, Metzger G. A 16-channel combined loop-dipole transceiver array for 7 Tesla body MRI. Magn Reson Med. 2017;77:884-894.
Weinberger O, Winter L, Dieringer MA, et al. Local multi-channel RF surface coil versus body RF coil transmission for cardiac magnetic resonance at 3 Tesla: Which configuration is winning the game? PLoS ONE. 2016;11:e0161863.
Niendorf T, Paul K, Oezerdem C, et al. W(h)ither human cardiac and body magnetic resonance at ultrahigh fields? Technical advances, practical considerations, applications, and clinical opportunities. NMR Biomed. 2016;29:1173-1197.
Kovacs H, Moskau D, Spraul M. Cryogenically cooled probes-A leap in NMR technology. Prog Nucl Magn Reson Spectrosc. 2005;46:131-155.
Baltes C, Radzwill N, Bosshard S, Marek D, Rudin M. Micro MRI of the mouse brain using a novel 400MHz cryogenic quadrature RF probe. NMR Biomed. 2009;22:834-842.
Niendorf T, Pohlmann A, Reimann HM, et al. Advancing cardiovascular, neurovascular, and renal magnetic resonance imaging in small rodents using cryogenic radiofrequency coil technology. Front Pharmacol. 2015;12:255.
Carvalho V, Hertanu A, Grélard A, et al. MRI assessment of multiple dipolar relaxation time (T1D) components in biological tissues interpreted with a generalized inhomogeneous magnetization transfer (ihMT) model. J Magn Reson. 2020;311:106668.
Axel L, Constantini J, Listerud J. Intensity correction in surface-coil MR imaging. Am J Roentgenol. 1987;148:418-420.
Wicks D, Barker G, Tofts P. Correction of intensity nonuniformity in MR images of any orientation. Magn Reson Imaging. 1993;11:183-196.
McVeigh E, Bronskill M, Henkelman R. Phase and sensitivity of receiver coils in magnetic resonance imaging. Med Phys. 1986;13:806-814.
Tincher M, Meyer C, Gupta R, Williams D. Polynomial modeling and reduction of RF body coil spatial inhomogeneity in MRI. IEEE Trans Med Imaging. 1993;12:361-365.
Meyer CR, Bland PH, Pipe J. Retrospective correction of intensity inhomogeneities in MRI. IEEE Trans Med Imaging. 1995;14:36-41.
Haselgrove J, Prammer M. An algorithm for compensation of surface-coil images for sensitivity of the surface coil. Magn Reson Imaging. 1986;4:469-472.
Moyher S, Vigneron D, Nelson S. Surface coil MR imaging of the human brain with an analytic reception profile correction. J Magn Reson Imaging. 1995;5:139-144.
Narayana P, Brey W, Kulkarny M, Sievenpiper C. Compensation for surface coil sensitivity variation in magnetic resonance imaging. Magn Reson Imaging. 1988;6:271-274.
Vaidya M, Collins C, Sodickson D, Brown R, Wiggins G, Lattanzi R. Dependence of B1+ and B1− field patterns of surface coils on the electrical properties of the sample and the MR operating frequency. Concept Magn Reson B. 2016;46B:25-40.
Keltner J, Carlson J, Roos M, Wong S, Wong T, Budinger T. Electromagnetic fields of surface coil in vivo NMR at high frequencies. Magn Reson Med. 1991;22:467-480.
Boulant N. T1 and T2 effects during radio-frequency pulses in spoiled gradient echo sequences. J Magn Reson. 2009;197:213-218.
Dieringer MA, Deimling M, Santoro D, et al. Rapid parametric mapping of the longitudinal relaxation time T1 using two-dimensional variable flip angle magnetic resonance imaging at 1.5 Tesla, 3 Tesla, and 7 Tesla. PLoS ONE. 2014;9:e91318.
Lommen J, Konstandin S, Krämer P, Schad L. Enhancing the quantification of tissue sodium content by MRI: Time-efficient sodium B1 mapping at clinical field strengths. NMR Biomed. 2015;29:129-136.
Waiczies S, Millward JM, Starke L, et al. Enhanced fluorine-19 MRI sensitivity using a cryogenic radiofrequency probe: Technical developments and ex vivo demonstration in a mouse model of neuroinflammation. Sci Rep. 2017;7:9808.
de Graaf R, Nicolay K. Adiabatic rf pulses: Applications to in vivo NMR. Magn Reson Med. 1998;40:690-696.
Baldanchani P, Pauly J, Spielman D. Designing adiabatic radio frequency pulses using the Shinnar-Le Roux algorithm. Magn Reson Med. 2010;64:843-851.
Zivkovic I, Teeuwisse W, Slobozhanyuk A, Nenasheva E, Webb A. High permittivity ceramics improve the transmit field and receive efficiency of a commercial extremity coil at 1.5 Tesla. J Magn Reson. 2019;299:59-65.
Brink W, van der Jagt A, Versluis M, Verbist B, Webb A. High permittivity dielectric pads improve high spatial resolution magnetic resonance imaging of the inner ear at 7 T. Invest Radiol. 2014;49:271-277.
O'Brien K, Magill A, Delacoste J, et al. Dielectric pads and low-B1+ adiabatic pulses: Complementary techniques to optimize structural T1w whole-brain MP2RAGE scans at 7 Tesla. J Magn Reson Imaging. 2014;40:804-812.
O'Reilly T, Webb A, Brink W. Practical improvements in the design of high permittivity pads for dielectric shimming in neuroimaging at 7 T. J Magn Reson. 2016;270:108-114.
van der Bergen B, van der Berg C, Bartels L, Lagendijk J. 7 T body MRI: B1 shimming with simultaneous SAR reduction. Phys Med Biol. 2007;52:5429-5441.
Winkler S, Rutt B. Practical methods for improving B1+ homogeneity in 3 Tesla breast imaging. J Magn Reson Imaging. 2015;41:992-999.
Yang QX, Mao W, Wang J, et al. Manipulation of image intensity distribution at 7.0 T: Passive RF shimming and focusing with dielectric materials. J Magn Reson Imaging. 2006;24:197-202.
Vernikouskaya I, Pochert A, Lindén M, Rasche V. Quantitative 19F MRI of perfluoro-15-crown-5-ether using uniformity correction of the spin excitation and signal reception. Magn Reson Mater Phys, Biol Med. 2019;32:25-36.
van Schie J, Lavini C, van Vliet L, Vos F. Feasibility of a fast method for B1-inhomogeneity correction for FSPGR sequences. Magn Reson Imaging. 2015;33:312-318.
Wang J, Qiu M, Constable R. In vivo method for correcting transmit/receive nonuniformities with phased array coils. Magn Reson Med. 2005;53:666-674.
Collewet G, Davenel A, Toussaint C, Akoka S. Correction of intensity nonuniformity in spin-echo T1-weighted images. Magn Reson Imaging. 2002;20:365-373.
Cooper M, Nguyen T, Spincemaille P, Prince M, Weinsaft J, Wang Y. Flip angle profile correction for T1 and T2 quantification with look-locker inversion recovery 2D SSFP imaging. Magn Reson Med. 2012;68:1579-1585.
Ma D, Coppo S, Chen Y, et al. Slice profile and B1 corrections in 2D magnetic resonance fingerprinting. Magn Reson Med. 2017;78:1781-1789.
Deoni S. Correction of main and transmit magnetic field (B0 and B1) inhomogeneity effects in multicomponent-driven equilibrium single-pulse observation of T1 and T2. Magn Reson Med. 2011;65:1021-1035.
Insko E, Bolinger L. Mapping of the radiofrequency field. J Magn Reson. 1993;103A:82-85.
Akoka S, Franconi F, Seguin F, Le Pape A. Radiofrequency map of an NMR coil by imaging. Magn Reson Imaging. 1993;11:437-441.
Morrell G. A phase-sensitive method of flip angle mapping. Magn Reson Med. 2008;60:889-894.
Yarnykh V. Actual flip-angle imaging in the pulsed steady state: A method for rapid three-dimensional mapping of the transmitted radiofrequency field. Magn Reson Med. 2007;57:192-200.
Wang J, Mao W, Qiu M, Smith M, Constable R. Factors influencing flip angle mapping in MRI: RF pulse shape, slice-select gradients, off-resonance excitation, and B0 inhomogeneities. Magn Reson Med. 2006;56:463-468.
Stollberger R, Wach P. Imaging of the active B1 field in vivo. Magn Reson Med. 1996;35:246-251.
Cunningham C, Pauly J, Nayak K. Saturated double-angle method for rapid B1+ mapping. Magn Reson Med. 2006;55:1326-1333.
Wang D, Heberlein K, LaConte S, Hu X. Inherent insensitivity to RF inhomogeneity in FLASH imaging. Magn Reson Med. 2004;52:927-931.
Bernstein M. Gradient echo. In: Bernstein M, King K, Zhou X, eds. Handbook of MRI Pulse Sequences. Burlington, MA, San Diego, CA and London: Elsevier Academic Press; 2004:579-605.
Hennig J, Nauerth A, Friedburg H. RARE imaging: A fast imaging method for clinical MR. Magn Reson Med. 1986;3:823-833.
Conturo T, Beth A, Arenstorf R, Price R. Simplified mathematical description of longitudinal recovery in multiple-echo sequences. Magn Reson Med. 1987;4:282-288.
Meara S, Barker G. Evolution of the longitudinal magnetization for pulse sequences using a fast spin-echo readout: Application to fluid-attenuated inversion-recovery and double inversion-recovery sequences. Magn Reson Med. 2005;54:241-245.
Breuer F, Blaimer M, Heidemann R, Mueller M, Griswold M, Jakob P. Controlled aliasing in parallel imaging results in higher acceleration (CAIPIRINHA) for multi-slice imaging. Magn Reson Med. 2005;53:684-691.
Skare S, Holdsworth S, Lilja A, Bammer R. Image domain propeller fast spin echo. Magn Reson Imaging. 2013;31:385-395.
Zhou X, Liang Z, Gewalt S, Cofer G, Lauterbur P, Johnson G. A fast spin echo technique with circular sampling. Magn Reson Med. 1998;39:23-27.
Altbach MI, Outwater EK, Trouard TP, et al. Radial fast spin-echo method for T2-weighted imaging and T2 mapping of the liver. J Magn Reson Imaging. 2002;16:179-189.
Busse RF, Brau ACS, Vu A, et al. Effects of refocusing flip angle modulation and view ordering in 3D fast spin echo. Magn Reson Med. 2008;60:640-649.
Reeder SB, Pineda AR, Wen Z, et al. Iterative decomposition of water and fat with echo asymmetry and least-squares estimation (IDEAL): Application with fast spin-echo imaging. Magn Reson Med. 2005;54:636-644.
Oshio K, Feinberg D. GRASE (Gradient- and Spin Echo) imaging: A novel fast MRI technique. Magn Reson Med. 1991;20:344-349.
Robson D, Gatehouse P, Bydder M, Bydder G. Magnetic resonance: An introduction to ultrashort TE (UTE) imaging. J Comput Assist Tomogr. 2003;27:825-846.
Mansfield P. Multi-planar image formation using NMR spin echoes. J Phys C: Solid State Phys. 1977;10:55-58.
Waiczies H, Lepore S, Drechsler S, et al. Visualizing brain inflammation with a shingled-leg radio-frequency head probe for 19F/1H MRI. Sci Rep. 2013;3:1280.
Polyfitn [computer program]. Version 1.3. File Exchange. Natick, MA: The MathWorks, Inc. 2016.
Ibrahim T, Hue Y, Tang L. Understanding and manipulating the RF fields at high field MRI. NMR Biomed. 2009;22:927-936.
Pauly J, Nishimura D, Macovski A. A k-space analysis of small-tip-angle excitation. J Magn Reson. 1989;81:43-56.
Manjón J, Coupé P, Martí-Bonmatí L, Collins D, Robles M. Adaptive non-local means denoising of MR images with spatially varying noise levels. J Magn Reson Imaging. 2010;31:192-203.
Price R, Axel L, Morgan T, et al. Quality assurance methods and phantoms for magnetic resonance imaging: Report of AAPM nuclear magnetic resonance task group No. 1. Med Phys. 1990;17:287-295
Feinberg DA, Mills CM, Posin JP, et al. Multiple spin-echo magnetic resonance imaging. Radiology. 1985;155:437-442.
Balezeau F, Eliat P, Cayamo A, Saint-Jalmes H. Mapping of low flip angles in magnetic resonance. Phys Med Biol. 2011;56:6635-6647.
Khalil AA, Mueller S, Foddis M, et al. Longitudinal 19F magnetic resonance imaging of brain oxygenation in a mouse model of vascular cognitive impairment using a cryogenic radiofrequency coil. Magn Reson Mater Phys, Biol Med. 2019;32:105-114.
Weigel M. Extended phase graphs: Dephasing, RF pulses, and echoes-Pure and simple. J Magn Reson Imaging. 2015;41:266-295.
Bloch F. Nuclear induction. Phys Rev. 1946;70:460-474.
Ma D, Gulani V, Seiberlich N, et al. Magnetic resonance fingerprinting. Nature. 2013;495:187-192.
Ratering D, Baltes C, Nordmeyer-Massner J, Marek D, Rudin M. Performance of a 200-MHz cryogenic RF probe designed for MRI and MRS of the murine brain. Magn Reson Med. 2008;59:1440-1447.