Bacteraemia after leadless pacemaker implantation.


Journal

Journal of cardiovascular electrophysiology
ISSN: 1540-8167
Titre abrégé: J Cardiovasc Electrophysiol
Pays: United States
ID NLM: 9010756

Informations de publication

Date de publication:
09 2020
Historique:
received: 27 04 2020
revised: 06 07 2020
accepted: 10 07 2020
pubmed: 16 7 2020
medline: 15 7 2021
entrez: 16 7 2020
Statut: ppublish

Résumé

Transvenous 3 permanent pacemaker-related infection is a severe condition associated with significant morbidity and mortality. Leadless pacemakers may be more resistant to bacterial seeding during bloodstream infection because of its small surface area and encapsulation in the right ventricle. This study reports the incidence and outcomes of bacteraemia in patients implanted with a Micra leadless pacemaker. We present We report a retrospective cohort study of 155 patients who underwent a Micra TPS implant procedure at the University Hospitals of Leuven between July 2015 and July 2019. We identified the patients who developed an episode of bacteraemia, proved by ≥2 positive blood cultures. Of the 155 patients, 15 patients presented an episode of bacteraemia at a median of 226 days (range: 3-1129) days after the implant procedure. Gram-positive species accounted for 73.3% (n = 11) of the bacteraemia including Staphylococcus (n = 5), Enterococcus (n = 3), and Streptococcus (n = 3). The source of infection was identified in nine patients (60%) including endocarditis in four patients, urinary tract in three patients, and skin in two patients. In our small cohort, no leadless pacemaker endocarditis was observed among patients with bacteraemia.

Sections du résumé

BACKGROUND
Transvenous 3 permanent pacemaker-related infection is a severe condition associated with significant morbidity and mortality. Leadless pacemakers may be more resistant to bacterial seeding during bloodstream infection because of its small surface area and encapsulation in the right ventricle. This study reports the incidence and outcomes of bacteraemia in patients implanted with a Micra leadless pacemaker. We present
METHODS
We report a retrospective cohort study of 155 patients who underwent a Micra TPS implant procedure at the University Hospitals of Leuven between July 2015 and July 2019. We identified the patients who developed an episode of bacteraemia, proved by ≥2 positive blood cultures.
RESULTS
Of the 155 patients, 15 patients presented an episode of bacteraemia at a median of 226 days (range: 3-1129) days after the implant procedure. Gram-positive species accounted for 73.3% (n = 11) of the bacteraemia including Staphylococcus (n = 5), Enterococcus (n = 3), and Streptococcus (n = 3). The source of infection was identified in nine patients (60%) including endocarditis in four patients, urinary tract in three patients, and skin in two patients.
CONCLUSION
In our small cohort, no leadless pacemaker endocarditis was observed among patients with bacteraemia.

Identifiants

pubmed: 32666611
doi: 10.1111/jce.14671
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

2440-2447

Informations de copyright

© 2020 Wiley Periodicals LLC.

Références

Habib G, Lancellotti P, Antunes MJ, et al. ESC Guidelines for the management of infective endocarditis: the Task Force for the Management of Infective Endocarditis of the European Society of Cardiology (ESC). Endorsed by: European Association for Cardio-Thoracic Surgery (EACTS), the European Association of Nuclear Medicine (EANM). Eur Heart J. 2015;36(44):3075-3128.
Bongiorni MG, Burri H, Deharo JC, et al. EHRA expert consensus statement on lead extraction: recommendations on definitions, endpoints, research trial design, and data collection requirements for clinical scientific studies and registries: endorsed by APHRS/HRS/LAHRS. Europace. 2018;20(7):1217.
Olsen T, Jorgensen OD, Nielsen JC, Thogersen AM, Philbert BT, Johansen JB. Incidence of device-related infection in 97 750 patients: clinical data from the complete Danish device-cohort (1982-2018). Eur Heart J. 2019;40(23):1862-1869.
Poole JE, Gleva MJ, Mela T, et al. Complication rates associated with pacemaker or implantable cardioverter-defibrillator generator replacements and upgrade procedures: results from the REPLACE registry. Circulation. 2010;122(16):1553-1561.
Chamis AL, Peterson GE, Cabell CH, et al. Staphylococcus aureus bacteremia in patients with permanent pacemakers or implantable cardioverter-defibrillators. Circulation. 2001;104(9):1029-1033.
Maskarinec SA, Thaden JT, Cyr DD, Ruffin F, Souli M, Fowler VG. The risk of cardiac device-related infection in bacteremic patients is species specific: results of a 12-year prospective cohort. Open Forum Infect Dis. 2017;4(3):ofx132.
Kusumoto FM, Schoenfeld MH, Wilkoff BL, et al. 2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm. 2017;14(12):e503-e551.
Sarrazin JF, Philippon F, Tessier M, et al. Usefulness of fluorine-18 positron emission tomography/computed tomography for identification of cardiovascular implantable electronic device infections. J Am Coll Cardiol. 2012;59(18):1616-1625.
Graziosi M, Nanni C, Lorenzini M, et al. Role of (1)(8)F-FDG PET/CT in the diagnosis of infective endocarditis in patients with an implanted cardiac device: a prospective study. Eur J Nucl Med Mol Imaging. 2014;41(8):1617-1623.
Bensimhon L, Lavergne T, Hugonnet F, et al. Whole body [(18) F]fluorodeoxyglucose positron emission tomography imaging for the diagnosis of pacemaker or implantable cardioverter defibrillator infection: a preliminary prospective study. Clin Microbiol Infect. 2011;17(6):836-844.
Reynolds D, Duray GZ, Omar R, et al. A leadless intracardiac transcatheter pacing system. N Engl J Med. 2016;374(6):533-541.
El-Chami MF, Al-Samadi F, Clementy N, et al. Updated performance of the Micra transcatheter pacemaker in the real-world setting: a comparison to the investigational study and a transvenous historical control. Heart Rhythm. 2018;15(12):1800-1807.
Satoh T, Fukui A, Katoh S, Matsui M. A leadless pacemaker which became encapsulated only two months after placement. Intern Med. 2018;57(20):3053-3054.
Kypta A, Blessberger H, Lichtenauer M, Steinwender C. Complete encapsulation of a leadless cardiac pacemaker. Clin Res Cardiol. 2016;105(1):94.
Brignole M, Auricchio A, Baron-Esquivias G, et al. ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: the Task Force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA). Eur Heart J. 2013;34(29):2281-2329.
Greenspon AJ, Rhim ES, Mark G, Desimone J, Ho RT. Lead-associated endocarditis: the important role of methicillin-resistant Staphylococcus aureus. Pacing Clin Electrophysiol. 2008;31(5):548-553.
Nagpal A, Baddour LM, Sohail MR. Microbiology and pathogenesis of cardiovascular implantable electronic device infections. Circ Arrhythm Electrophysiol. 2012;5(2):433-441.
El-Chami MF, Johansen JB, Zaidi A, et al. Leadless pacemaker implant in patients with pre-existing infections: results from the Micra postapproval registry. J Cardiovasc Electrophysiol. 2019;30(4):569-574.
El-Chami MF, Soejima K, Piccini JP, et al. Incidence and outcomes of systemic infections in patients with leadless pacemakers: data from the Micra IDE study. Pacing Clin Electrophysiol. 2019;42(8):1105-1110.
Erba PA, Lancellotti P, Vilacosta I, et al. Recommendations on nuclear and multimodality imaging in IE and CIED infections. Eur J Nucl Med Mol Imaging. 2018;45(10):1795-1815.
El-Chami MF, Mayotte J, Bonner M, Holbrook R, Stromberg K, Sohail MR. Reduced bacterial adhesion with parylene coating: potential implications for Micra transcatheter pacemakers. J Cardiovasc Electrophysiol. 2020;31:712-717.
Hill EE, Herijgers P, Claus P, Vanderschueren S, Herregods MC, Peetermans WE. Infective endocarditis: changing epidemiology and predictors of 6-month mortality: a prospective cohort study. Eur Heart J. 2007;28(2):196-203.
Polyzos KA, Konstantelias AA, Falagas ME. Risk factors for cardiac implantable electronic device infection: a systematic review and meta-analysis. Europace. 2015;17(5):767-777.

Auteurs

Christophe Garweg (C)

Department of Cardiology, University Hospitals Leuven, Leuven, Belgium.
Department of Cardiovascular Sciences, University of Leuven, Leuven, Belgium.

Bert Vandenberk (B)

Department of Cardiology, University Hospitals Leuven, Leuven, Belgium.

Sander Jentjens (S)

Department of Nuclear Medicine, University Hospitals Leuven, Leuven, Belgium.

Stefaan Foulon (S)

Department of Cardiology, University Hospitals Leuven, Leuven, Belgium.

Patrick Hermans (P)

Department of Cardiology, University Hospitals Leuven, Leuven, Belgium.

Patricia Poels (P)

Department of Cardiology, University Hospitals Leuven, Leuven, Belgium.

Peter Haemers (P)

Department of Cardiology, University Hospitals Leuven, Leuven, Belgium.
Department of Cardiovascular Sciences, University of Leuven, Leuven, Belgium.

Joris Ector (J)

Department of Cardiology, University Hospitals Leuven, Leuven, Belgium.
Department of Cardiovascular Sciences, University of Leuven, Leuven, Belgium.

Rik Willems (R)

Department of Cardiology, University Hospitals Leuven, Leuven, Belgium.
Department of Cardiovascular Sciences, University of Leuven, Leuven, Belgium.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH