Precise Signals with a High-Density Grid Mapping Catheter Are Useful for an Entrainment Study.
Entrainment mapping
Post MAZE surgery
Journal
International heart journal
ISSN: 1349-3299
Titre abrégé: Int Heart J
Pays: Japan
ID NLM: 101244240
Informations de publication
Date de publication:
30 Jul 2020
30 Jul 2020
Historique:
pubmed:
21
7
2020
medline:
13
8
2020
entrez:
21
7
2020
Statut:
ppublish
Résumé
Complex atrial tachycardias (ATs) after catheter ablation or a MAZE procedure is sometimes difficult to determine the circuits of the tachycardia. A high-density, grid-shapes mapping catheter has been launched, which can be useful for detecting the detail circuits of tachycardias on three-dimensional mapping systems. The signal quality is also important for performing electrophysiological studies (EPSs), such as entrainment mapping, to identify the circuit. This unique mapping catheter has 1 mm electrodes on 2.5 Fr shafts, which improve the signal quality. The high-quality intracardiac electrograms facilitate differentiating small critical potentials, which allows us to perform detailed entrainment mapping in targeted narrow areas. Here, we describe a patient with a perimetral AT with epi-endocardium breakthrough after a MAZE surgery and catheter ablation, which was treated successfully along with detailed entrainment mapping using the HD Grid. This catheter with high-quality signals could be a significant diagnostic tool for a classic EPS as well as for the construction of 3D mapping.
Types de publication
Case Reports
Journal Article
Langues
eng
Sous-ensembles de citation
IM