Use of pulse contour technology for continuous blood pressure monitoring in pediatric patients.


Journal

Blood pressure monitoring
ISSN: 1473-5725
Titre abrégé: Blood Press Monit
Pays: England
ID NLM: 9606438

Informations de publication

Date de publication:
Oct 2020
Historique:
pubmed: 24 7 2020
medline: 5 1 2021
entrez: 24 7 2020
Statut: ppublish

Résumé

This study evaluates the accuracy of continuous blood pressure monitoring using pulse contour technology with the ClearSight monitoring device, a noninvasive alternative to placing an invasive arterial line, in pediatric patients. Children younger than 18 years admitted to a pediatric ICU, who required an arterial line, and fit into the ClearSight finger cuff were included. Blood pressure measurement for systolic, diastolic, and mean arterial pressures (MAP) obtained by the ClearSight device were compared with those obtained with the intra-arterial catheter as well as automated cuff measurements using the mixed-effects model. Analysis was conducted for entire cohort, and measurements obtained with and without vasopressor use. There were 213 measurements from 10 patients. There was a statistically significant difference in systolic blood pressure when comparing arterial line and ClearSight systolic and diastolic measurements between the two methods (P < 0.001). There was no statistical difference between arterial MAP and ClearSight MAP (P = 0.957). Results were similar when ClearSight measurements were compared with automated cuff measurements. Both the vasopressor use and nonvasopressor use groups showed a statistically significant difference between arterial and ClearSight measurements for systolic and diastolic pressures, but not for the MAP. Measurements of MAP obtained by the ClearSight device were almost identical to those obtained by the intra-arterial catheter. Although there was a difference in systolic blood pressures between the two methods, in those patients receiving inotropic support, the difference was within the range of what is considered acceptable in validating blood pressure devices.

Identifiants

pubmed: 32701567
doi: 10.1097/MBP.0000000000000458
pii: 00126097-202010000-00009
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

278-284

Références

Mignini MA, Piacentini E, Dubin A. Peripheral arterial blood pressure monitoring adequately tracks central arterial blood pressure in critically ill patients: an observational study. Crit Care. 2006; 10:R43
Hebal F, Sparks HT, Rychlik KL, Bone M, Tran S, Barsness KA. Pediatric arterial catheters: complications and associated risk factors. J Pediatr Surg. 2018; 53:794–797
King MA, Garrison MM, Vavilala MS, Zimmerman JJ, Rivara FP. Complications associated with arterial catheterization in children. Pediatr Crit Care Med. 2008; 9:367–371
Meidert AS, Saugel B. Techniques for non-invasive monitoring of arterial blood pressure. Front Med (Lausanne). 2017; 4:231
Bogert LW, Wesseling KH, Schraa O, Van Lieshout EJ, de Mol BA, van Goudoever J, et al. Pulse contour cardiac output derived from non-invasive arterial pressure in cardiovascular disease. Anaesthesia. 2010; 65:1119–1125
Bubenek-Turconi SI, Craciun M, Miclea I, Perel A. Noninvasive continuous cardiac output by the Nexfin before and after preload-modifying maneuvers. Anesth Analg. 2013; 117:366–372
Fischer MO, Avram R, Cârjaliu I, Massetti M, Gérard JL, Hanouz JL, et al. Non-invasive continuous arterial pressure and cardiac index monitoring with Nexfin after cardiac surgery. Br J Anaesth. 2012; 109:514–521
Martina JR, Westerhof BE, van Goudoever J, de Beaumont EM, Truijen J, Kim YS, et al. Noninvasive continuous arterial blood pressure monitoring with Nexfin®. Anesthesiology. 2012; 116:1092–1103
Penaz J. Photoelectric measurement of blood pressure, volume and flow in the finger. 1973Digest of the 10th International Conference on Medical and Biomedical EngineeringDresden, Germany
Wesseling KH, de Wit B, van der Hoeven GMA, van Goudoever J, Settels JJ. Physiocal, calibrating finger vascular physiology for finapres. Homeostasis. 1995; 36:76–82
Truijen J, van Lieshout JJ, Wesselink WA, Westerhof BE. Noninvasive continuous hemodynamic monitoring. J Clin Monit Comput. 2012; 26:267–278
Maki DG, Kluger DM, Crnich CJ. The risk of bloodstream infection in adults with different intravascular devices: a systematic review of 200 published prospective studies. Mayo Clin Proc. 2006; 81:1159–1171
O’Horo JC, Maki DG, Krupp AE, Safdar N. Arterial catheters as a source of bloodstream infection: a systematic review and meta-analysis. Crit Care Med. 2014; 42:1334–1339
Gershengorn HB, Wunsch H, Scales DC, Zarychanski R, Rubenfeld G, Garland A. Association between arterial catheter use and hospital mortality in intensive care units. JAMA Intern Med. 2014; 174:1746–1754
Antonova ML. Recent patents on accuracy of blood pressure measurement. Recent Patents Biomed Eng. 2013; 6:58–73
Creagh-Brown B. 2004
Stover JF, Stocker R, Lenherr R, Neff TA, Cottini SR, Zoller B, Béchir M. Noninvasive cardiac output and blood pressure monitoring cannot replace an invasive monitoring system in critically ill patients. BMC Anesthesiol. 2009; 9:6
Critchley LA. Pulse contour analysis: is it able to reliably detect changes in cardiac output in the haemodynamically unstable patient? Crit Care. 2011; 15:106
Hadian M, Kim HK, Severyn DA, Pinsky MR. Cross-comparison of cardiac output trending accuracy of LiDCO, PiCCO, FloTrac and pulmonary artery catheters. Crit Care. 2010; 14:R212
Stergiou GS, Alpert B, Mieke S, Asmar R, Atkins N, Eckert S, et al. A universal standard for the validation of blood pressure measuring devices: Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Collaboration Statement. J Hypertens. 2018; 36:472–478
Marik PE. Obituary: pulmonary artery catheter 1970 to 2013. Ann Intensive Care. 2013; 3:38

Auteurs

Hussam Alharash (H)

Department of Pediatric Critical Care Nemours/AI duPont Hospital For Children.

Md Jobayer Hossain (MJ)

Department of Biomedical Research, Nemours/Alfred I. duPont Hospital for Children, Wilmington, Delaware.

Abhishek Bhattacharjee (A)

Department of Biomedical Research, Nemours Children's Health System, Jacksonville, Florida.

Yosef Levenbrown (Y)

Department of Pediatric Critical Care Nemours/AI duPont Hospital For Children.
Department of Pediatrics, Sidney Kimmel Medical School of Thomas Jefferson University, Philadelphia, Pennsylvania, USA.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH