Comparison of free breathing 3D mDIXON with single breath-hold 3D inversion recovery sequences for the assessment of Late Gadolinium Enhancement.
DIXON based fat-supression
Magnetic resonance imaging
Three-dimensional Image
Journal
European journal of radiology
ISSN: 1872-7727
Titre abrégé: Eur J Radiol
Pays: Ireland
ID NLM: 8106411
Informations de publication
Date de publication:
Jan 2021
Jan 2021
Historique:
received:
01
04
2020
revised:
30
09
2020
accepted:
14
11
2020
pubmed:
12
12
2020
medline:
15
4
2021
entrez:
11
12
2020
Statut:
ppublish
Résumé
To evaluate the technical and diagnostic performance of three dimensional (3D) mDIXON versus 3D inversion recovery (3D VIAB) and 3D spectral presaturation with inversion recovery (3D SPIR) late gadolinium enhancement (LGE) sequences. A total of 78 patients (50 males and 28 females, age 49 ± 18 years) with 1.5 T CMR examination including three different 3D LGE sequences (3D mDIXON, 3D VIAB, and 3D SPIR) were evaluated for technical and diagnostic performance by two readers. Qualitative scores and quantitative signal and contrast-to-noise ratios were compared among sequences. Qualitative comparisons were made using Friedman and Wilcoxon signed rank tests. Quantitative comparisons were made using one way ANOVA. Reader agreements were tested using Cohen's Kappa. Any p-value <0.05 was significant. 19 out of 78 patients (24 %) were excluded due to poor (grade 4) image quality and 29 patients were excluded due to absence of LGE. For the remaining 30 patients, free breathing 3D mDIXON showed higher confidence in diagnosis of subepicardial LGE (p-value < 0.05). 3D mDIXON outperformed 3D SPIR in both visualization of LGE (p = 0.02) and quality of fat suppression (p = 0.001). Nevertheless, 3D mDIXON showed lower image quality compared to the other two sequences. Free breathing 3D mDIXON is a diagnostic problem-solving tool, especially when making a diagnosis of subepicardial enhancement and/or fat suppression is needed, owing to its high spatial resolution and robust fat suppression. Choice of 3D LGE sequence should be based on patient's breath-hold ability, diagnostic needs, and institutional availability considering the strengths and limitations of each sequence.
Identifiants
pubmed: 33307461
pii: S0720-048X(20)30617-3
doi: 10.1016/j.ejrad.2020.109427
pii:
doi:
Substances chimiques
Contrast Media
0
Gadolinium
AU0V1LM3JT
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
109427Informations de copyright
Copyright © 2020 Elsevier B.V. All rights reserved.