Optimization of Alanine Measurements for Fast and Accurate Dosimetry in FLASH Radiation Therapy.


Journal

Radiation research
ISSN: 1938-5404
Titre abrégé: Radiat Res
Pays: United States
ID NLM: 0401245

Informations de publication

Date de publication:
01 Dec 2020
Historique:
received: 19 11 2019
accepted: 17 02 2020
entrez: 21 12 2020
pubmed: 22 12 2020
medline: 9 3 2021
Statut: ppublish

Résumé

FLASH radiation therapy (FLASH-RT) reference dosimetry to obtain traceability, repeatability and stability of irradiations cannot be performed with conventional dosimetric methods, such as monitor chambers or ionization chambers. Until now, only passive dosimeters have provided the necessary dosimetric data. Alanine dosimetry is accurate; however, to be used for FLASH-RT in biological experiments and for clinical transfer to humans, the reading time needs to be reduced, while preserving a maximum deviation to the reference of ±2%. Optimization of alanine dosimetry was based on the acquisition of electron paramagnetic resonance (EPR) spectra with a Bruker spectrometer. Reading parameters such as the conversion time, the number of scans, the time constant, the microwave power and the modulation amplitude of the magnetic field were optimized as a trade-off between the signal-to-noise ratio (SNR) and the reading time of one measurement using the reference 10.1 Gy alanine pellet. After optimizing the parameters, we compared the doses measured with alanine pellets up to 100 Gy with the reference doses, and then determined the number of measurements necessary to get a difference lower than ±2%. A low-dose alanine pellet of 4.9 Gy was also measured to evaluate the quality of the optimization for doses lower than 10 Gy. The optimization of the Bruker default parameters made it possible to reduce the reading time for one measurement from 5.6 to 2.6 min. That reduction was not at the cost of the SNR because it was kept comparable to the default parameters. Three measurements were enough to obtain a maximum dose deviation to the reference of 1.8% for the range of 10-100 Gy. The total reading time for the three measurements was 7.8 min (3 × 2.6 min). For lower doses such as 4.9 Gy, three measurements led to a deviation greater than 5%. By increasing the number of measurements to five, the average difference to the reference dose was reduced to less than 5% with a total reading time increased to 13.0 min. For doses between 10 Gy and 100 Gy, the optimized acquisition parameters made it possible to keep the average differences between the reference and the measured doses below ±2%, for a reading time of 7.8 min. This enabled an accurate and fast dose determination for biological preparations as part of FLASH-beam irradiations.

Identifiants

pubmed: 33348370
pii: 436365
doi: 10.1667/RR15568.1
doi:

Substances chimiques

Alanine OF5P57N2ZX

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

573-579

Informations de copyright

©2020 by Radiation Research Society. All rights of reproduction in any form reserved.

Auteurs

Maude Gondré (M)

Institute of Radiation Physics, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland.

Patrik Gonçalves Jorge (PG)

Institute of Radiation Physics, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland.

Marie-Catherine Vozenin (MC)

Radio-Oncology Department, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland.

Jean Bourhis (J)

Radio-Oncology Department, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland.

François Bochud (F)

Institute of Radiation Physics, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland.

Claude Bailat (C)

Institute of Radiation Physics, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland.

Raphaël Moeckli (R)

Institute of Radiation Physics, Lausanne University Hospital and University of Lausanne, Lausanne, Switzerland.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH