High-Performance Liquid Chromatography Method for Simultaneous Determination of Guaifenesin, Salbutamol Sulfate and Guaifenesin Impurity (Guaiacol).
Journal
Journal of chromatographic science
ISSN: 1945-239X
Titre abrégé: J Chromatogr Sci
Pays: United States
ID NLM: 0173225
Informations de publication
Date de publication:
21 Apr 2021
21 Apr 2021
Historique:
received:
13
01
2019
pubmed:
13
1
2021
medline:
29
6
2021
entrez:
12
1
2021
Statut:
ppublish
Résumé
A simple and reliable HPLC method is developed for isocratic separation of a ternary mixture of Salbutamol Sulfate (SAL), Guaifenesin (GUI) and its impurity Guaiacol (GUA) either in pure powder or in pharmaceutical formulation. Chromatographic separation was applied on a Hypersil GOLD CN column with a mobile phase consisting of 0.05 M KH2PO4 (containing 0.1% triethylamine, pH adjusted to 3.7 by phosphoric acid): methanol (60: 40 by volume) using 0.6 mL min-1 flow rate and detection of peaks at 275 nm at 25°C with run time around 6 min. The calibration plots were linear over the concentration ranges of 0.5-20, 0.5-30 and 0.1-10 μg mL-1 for SAL, GUI and GUA, respectively. ICH guidelines were used for the validation of presented method. The obtained results were compared with the results obtained from the reported HPLC method and no significant difference was found regarding accuracy and precision.
Identifiants
pubmed: 33434919
pii: 6090331
doi: 10.1093/chromsci/bmaa128
doi:
Substances chimiques
Guaifenesin
495W7451VQ
Guaiacol
6JKA7MAH9C
Albuterol
QF8SVZ843E
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
419-424Informations de copyright
© The Author(s) 2021. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.