Synchrotron-based micro computed tomography investigation of the implant-abutment fatigue-induced microgap changes.

Dental implant-abutment connection Dental implant-abutment design Dental implant-abutment interface Dental implants Morse taper

Journal

Journal of the mechanical behavior of biomedical materials
ISSN: 1878-0180
Titre abrégé: J Mech Behav Biomed Mater
Pays: Netherlands
ID NLM: 101322406

Informations de publication

Date de publication:
04 2021
Historique:
received: 03 11 2020
revised: 07 01 2021
accepted: 08 01 2021
pubmed: 31 1 2021
medline: 15 5 2021
entrez: 30 1 2021
Statut: ppublish

Résumé

This study evaluates the effect of dynamic-loading on the microgap of the IAC when different supratructure heights are applied. Forty-eight dental implants (24 each of butt-joint (H) and internal-conical connections (C)) were tested in this study. Each group was further divided into three groups (n = 8) according to the applied suprastructure height (H1, C1: 10 mm, H2, C2: 14 mm and H3, C3: 18 mm). All specimens were subjected to cyclic loading in a chewing-simulator with a load of 98 N for 5 × 10 SRμCT revealed an internal microgap range between 0.26 μm and 0.5 μm in the group C, whereas the group H exhibited a microgap range between 0.26 μm and 0.47 μm prior to loading. After chewing simulation, a smaller microgap size in all groups was detected ranging from 0.11 μm to 0.26 μm (group C: 0.11μm-0.26 μm; group H: 0.21μm-0.25 μm). The LM investigation showed mean microgap values at the outer IAC junction before loading from 5.8 μm to 11.3 μm and from 3.9 μm to 7.2 μm after loading. All specimens exhibited a vertical intrusion displacement of the abutment. Regardless of the crown height, the microgap between the abutment and implant systematically decreased after loading in both butt-joint and internal-conical connections.

Identifiants

pubmed: 33516129
pii: S1751-6161(21)00022-9
doi: 10.1016/j.jmbbm.2021.104330
pii:
doi:

Substances chimiques

Dental Implants 0

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

104330

Informations de copyright

Copyright © 2021 Elsevier Ltd. All rights reserved.

Auteurs

Aimen Bagegni (A)

Medical Center - University of Freiburg, Center for Dental Medicine, Department of Prosthetic Dentistry, Faculty of Medicine - University of Freiburg, Freiburg, Germany. Electronic address: aimen.awad.bagegni@uniklinik-freiburg.de.

Simon Zabler (S)

Lehrstuhl für Röntgenmikroskopie der Universität Würzburg, Josef-Martin-Weg 63 Hubland Nord, 97074 Würzburg, Germany.

Katja Nelson (K)

Medical Center - University of Freiburg, Center for Dental Medicine, Department of Oral and Craniomaxillofacial Surgery, Faculty of Medicine - University of Freiburg, Freiburg, Germany.

Alexander Rack (A)

ESRF-The European Synchrotron, 71 Avenue des Martyrs, 38000 Grenoble, France.

Benedikt C Spies (BC)

Medical Center - University of Freiburg, Center for Dental Medicine, Department of Prosthetic Dentistry, Faculty of Medicine - University of Freiburg, Freiburg, Germany.

Kirstin Vach (K)

Medical Center - University of Freiburg, Institute of Medical Biometry and Statistics, Faculty of Medicine - University of Freiburg, Freiburg, Germany.

Ralf Kohal (R)

Medical Center - University of Freiburg, Center for Dental Medicine, Department of Prosthetic Dentistry, Faculty of Medicine - University of Freiburg, Freiburg, Germany.

Articles similaires

Hemiarthroplasty in young patients.

Hazimah Mahmud, Dong Wang, Andra Topan-Rat et al.
1.00
Humans Male Hemiarthroplasty Middle Aged Aged
Silicon Dioxide Water Hot Temperature Compressive Strength X-Ray Diffraction
Nitriles Tensile Strength Materials Testing Gloves, Protective Product Packaging
Calcium Carbonate Sand Powders Construction Materials Materials Testing

Classifications MeSH