Temperature and current flow effects of different electrode placement in shoulder capacitive-resistive electric transfer applications: a cadaveric study.
CRet
Cadaver
Glenohumeral capsule
Physical therapy
Shoulder
Supraspinatus tendon
Journal
BMC musculoskeletal disorders
ISSN: 1471-2474
Titre abrégé: BMC Musculoskelet Disord
Pays: England
ID NLM: 100968565
Informations de publication
Date de publication:
04 Feb 2021
04 Feb 2021
Historique:
received:
03
11
2020
accepted:
14
12
2020
entrez:
5
2
2021
pubmed:
6
2
2021
medline:
15
5
2021
Statut:
epublish
Résumé
Impingement syndrome is currently estimated to represent 60% of all shoulder pain disorders. Capacitive-Resistive electric transfer therapy is aimed to provoke temperature and current flow changes in superficial and deep tissues. This in vitro study has evaluated the variation of temperature and current flow in the shoulder tissues during two different areas of application of the movable capacitive-resistive electric transfer electrode. A cross-sectional study designed, five fresh cryopreserved cadavers (10 shoulders) were included in this study. Four interventions (capacitive and resistive modes; low- and high-power) were performed for 5 min each by a diathermy "T-Plus" device in two shoulder regions: postero-superior and antero-lateral. Supraspinatus tendon, glenohumeral capsule and superficial temperatures were recorded at 1-min intervals and 5 min after treatment. A statistically significant difference was found only for the superficial area and time interaction, with high power-resistive application at the postero-superior shoulder area (P< 0.035). All the applications showed a 5 min after treatment temperature increase compared with the basal data, in all the application points. Superficial temperature in the high power-resistive application showed the greatest percent increase (42.93% ± 22.58), followed by the temperature in the tendon area with the same high power-resistive application (22.97% ± 14.70). The high power-resistive application showed the greatest percent of temperature increase in the applications, reaching 65.9% ± 22.96 at 5-min at the superficial level, and 32% ± 24.25 at 4-min at the level of the supraspinatus tendon. At the capsule level, high power-resistive was also the application that showed the greatest percent of increase, with 21.52% ± 16.16. The application with the lowest percent of temperature increase was the low power-capacitive, with a mean value of 4.86% at supraspinatus tendon level and 7.47% at capsular level. The shoulder postero-superior or antero-lateral areas of application of capacitive-resistive electric transfer did not cause statistically significant differences in the temperature changes in either supraspinatus tendon or glenohumeral capsule tissues in cadaveric samples. The high power-resistive application in the postero-superior area significantly increased superficial temperature compared with the same application in the antero-lateral position area.
Sections du résumé
BACKGROUND
BACKGROUND
Impingement syndrome is currently estimated to represent 60% of all shoulder pain disorders. Capacitive-Resistive electric transfer therapy is aimed to provoke temperature and current flow changes in superficial and deep tissues. This in vitro study has evaluated the variation of temperature and current flow in the shoulder tissues during two different areas of application of the movable capacitive-resistive electric transfer electrode.
METHODS
METHODS
A cross-sectional study designed, five fresh cryopreserved cadavers (10 shoulders) were included in this study. Four interventions (capacitive and resistive modes; low- and high-power) were performed for 5 min each by a diathermy "T-Plus" device in two shoulder regions: postero-superior and antero-lateral. Supraspinatus tendon, glenohumeral capsule and superficial temperatures were recorded at 1-min intervals and 5 min after treatment.
RESULTS
RESULTS
A statistically significant difference was found only for the superficial area and time interaction, with high power-resistive application at the postero-superior shoulder area (P< 0.035). All the applications showed a 5 min after treatment temperature increase compared with the basal data, in all the application points. Superficial temperature in the high power-resistive application showed the greatest percent increase (42.93% ± 22.58), followed by the temperature in the tendon area with the same high power-resistive application (22.97% ± 14.70). The high power-resistive application showed the greatest percent of temperature increase in the applications, reaching 65.9% ± 22.96 at 5-min at the superficial level, and 32% ± 24.25 at 4-min at the level of the supraspinatus tendon. At the capsule level, high power-resistive was also the application that showed the greatest percent of increase, with 21.52% ± 16.16. The application with the lowest percent of temperature increase was the low power-capacitive, with a mean value of 4.86% at supraspinatus tendon level and 7.47% at capsular level.
CONCLUSION
CONCLUSIONS
The shoulder postero-superior or antero-lateral areas of application of capacitive-resistive electric transfer did not cause statistically significant differences in the temperature changes in either supraspinatus tendon or glenohumeral capsule tissues in cadaveric samples. The high power-resistive application in the postero-superior area significantly increased superficial temperature compared with the same application in the antero-lateral position area.
Identifiants
pubmed: 33541324
doi: 10.1186/s12891-020-03918-7
pii: 10.1186/s12891-020-03918-7
pmc: PMC7860630
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
139Références
J Orthop Res. 2015 Jun;33(6):793-9
pubmed: 25640030
PLoS One. 2014 Jan 08;9(1):e84636
pubmed: 24416255
Phys Ther Sport. 2018 Jul;32:227-234
pubmed: 29870922
Int J Hyperthermia. 2011;27(6):591-603
pubmed: 21846195
BMC Musculoskelet Disord. 2020 Jan 20;21(1):46
pubmed: 31959172
J Altern Complement Med. 2020 Feb;26(2):147-153
pubmed: 31580698
Dtsch Arztebl Int. 2017 Nov 10;114(45):765-776
pubmed: 29202926
Acta Biomed. 2018 Jan 19;89(1-S):186-196
pubmed: 29350647
Acta Biomater. 2017 Nov;63:18-36
pubmed: 28867648
Int J Hyperthermia. 2017 Sep;33(6):696-702
pubmed: 28139939
Orthop Res Rev. 2018 Oct 23;10:83-91
pubmed: 30774463
J Sport Rehabil. 2019 Jul 1;28(5):459-463
pubmed: 29405818
Cell Physiol Biochem. 2014;34(5):1741-55
pubmed: 25427571
Int J Hyperthermia. 2012;28(1):9-18
pubmed: 22235780
J Phys Ther Sci. 2018 May;30(5):719-725
pubmed: 29765189
Temperature (Austin). 2017 Aug 4;4(3):227-257
pubmed: 28944269
Curr Drug Discov Technol. 2015;12(4):218-24
pubmed: 26778084
Shoulder Elbow. 2017 Apr;9(2):75-84
pubmed: 28405218