On the dependence of quantitative diffusion-weighted imaging on scanner system characteristics and acquisition parameters: A large multicenter and multiparametric phantom study with unsupervised clustering analysis.
Apparent diffusion coefficient
Diffusion-weighted imaging
Magnetic resonance imaging
Quality assurance
Journal
Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)
ISSN: 1724-191X
Titre abrégé: Phys Med
Pays: Italy
ID NLM: 9302888
Informations de publication
Date de publication:
May 2021
May 2021
Historique:
received:
21
01
2021
revised:
31
03
2021
accepted:
23
04
2021
pubmed:
16
5
2021
medline:
29
6
2021
entrez:
15
5
2021
Statut:
ppublish
Résumé
The purpose of this multicenter phantom study was to exploit an innovative approach, based on an extensive acquisition protocol and unsupervised clustering analysis, in order to assess any potential bias in apparent diffusion coefficient (ADC) estimation due to different scanner characteristics. Moreover, we aimed at assessing, for the first time, any effect of acquisition plan/phase encoding direction on ADC estimation. Water phantom acquisitions were carried out on 39 scanners. DWI acquisitions (b-value = 0-200-400-600-800-1000 s/mm No significant dependence of in-center ADC values on acquisition plan/phase encoding direction was found. Ward unsupervised clustering analysis showed 3 distinct clusters of scanners and an association between Δ-values and manufacturer/model, whereas no association between Δ-values and maximum gradient strength, slew rate or static magnetic field strength was revealed. Several acquisition setups showed significant differences among groups, indicating the introduction of different biases in ADC estimation. Unsupervised clustering analysis of DWI data, obtained from several scanners using an extensive acquisition protocol, allows to reveal an association between measured ADC values and manufacturer/model of scanner, as well as to identify suboptimal DWI acquisition setups for accurate ADC estimation.
Identifiants
pubmed: 33991807
pii: S1120-1797(21)00177-0
doi: 10.1016/j.ejmp.2021.04.020
pii:
doi:
Types de publication
Journal Article
Multicenter Study
Langues
eng
Sous-ensembles de citation
IM
Pagination
98-106Informations de copyright
Copyright © 2021 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.