Predicting the Postoperative Addition Power of a Multifocal Intraocular Lens at the Spectacle Plane.
Journal
Journal of refractive surgery (Thorofare, N.J. : 1995)
ISSN: 1938-2391
Titre abrégé: J Refract Surg
Pays: United States
ID NLM: 9505927
Informations de publication
Date de publication:
May 2021
May 2021
Historique:
entrez:
28
5
2021
pubmed:
29
5
2021
medline:
19
8
2021
Statut:
ppublish
Résumé
To establish a simple clinical method of predicting addition power achieved with a multifocal intraocular lens (IOL). In this prospective cohort study, 41 patients were bilaterally implanted with the Bi-Flex MY multifocal IOL (Medicontur) with +3.50 diopters (D) near addition power. Monocular defocus curves were plotted for each patient and effective addition power was calculated as the dioptric difference between the distance and near inflection points of the defocus curve. Six biometry formulas (Haigis, Holladay, SRK/T, Hill RBF, Barrett Universal II, and Holladay 2) were used to predict the addition power at the spectacle plane. Mean effective addition power was 2.60 ± 0.29 D, with significant ( The study demonstrates that the effective addition power can be predicted using the proposed simple clinical method derived using the Barrett Universal II formula. The proposed technique may have significant clinical value in screening for patients where ocular biometry may lead to aberrant addition power.