The STRYDE limb lengthening nail is susceptible to mechanically assisted crevice corrosion: an analysis of 23 retrieved implants.
Journal
Acta orthopaedica
ISSN: 1745-3682
Titre abrégé: Acta Orthop
Pays: Sweden
ID NLM: 101231512
Informations de publication
Date de publication:
Oct 2021
Oct 2021
Historique:
pubmed:
10
6
2021
medline:
16
11
2021
entrez:
9
6
2021
Statut:
ppublish
Résumé
Background and purpose - We noted several adverse events in patients in whom the first version of the STRYDE limb-lengthening nail (NuVasive Specialized Orthopaedics, San Diego, CA) had been implanted. Pain, osteolysis, periosteal reactions, and cortical hypertrophy at the nail junction were noted. Here, we present the analysis of 23 retrieved STRYDE implants.Materials and methods - We undertook visual inspection of the retrieved nails and screws, mechanical evaluation of the junction, micro-CT analyses, microscopic inspection of the bushing, screws, screw holes, and separated parts of the implants. Positive material identification (PMI) and energy-dispersive X-ray spectroscopy (EDS) were used to analyze the chemical composition. The hardness of the material was also investigated.Results - 20/23 retrieved nails had visible signs of corrosion, i.e., discoloration at the telescopic junction. Micro-CT verified corrosion attacks in 12/12 scanned bushings. Corrosion, predominantly mechanically assisted crevice corrosion, was observed at the locking screws and screw holes in 20/23 nails. Biological material inside the nail was observed in addition to oozing from the junction of 2 nails during hardware removal, which was experimentally reproducible. Notably, the mechanical construction of the bushing changed from PRECICE P2 to STRYDE nails.Interpretation - STRYDE nails are not hermetically sealed, and liquid can pass the bushing. Biodur 108 itself is corrosion resistant; however, mechanically assisted crevice corrosion of the bushing, locking screws, and screw holes may be aggravated due to manufacturing aiming for increased strength and hardness of the alloy.Observing several adverse events, we recently published a nationwide cross-sectional analysis of all 30 STRYDE limb- lengthening nails (NuVasive, Specialized Orthopedics, San Diego, CA) that were implanted in Denmark (Rölfing et al. 2021a). 27/30 STRYDE nails have now been removed and we present data from metallurgical analysis of 23 of the retrieved implants.
Identifiants
pubmed: 34102950
doi: 10.1080/17453674.2021.1927506
pmc: PMC8519523
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
621-627Références
J Pediatr Orthop. 2021 Jan;41(1):e44-e49
pubmed: 32947442
Bone Joint J. 2019 Sep;101-B(9):1168-1176
pubmed: 31474141
Eur Spine J. 2021 Jul;30(7):1905-1911
pubmed: 33544222
Orthopade. 2019 Jun;48(6):477-485
pubmed: 31172228
World Neurosurg. 2019 May;125:e593-e601
pubmed: 30716487
Bone Joint J. 2021 Jun;103-B(6):1168-1172
pubmed: 34058877
Orthop Traumatol Surg Res. 2021 May 18;:102958
pubmed: 33965599
Bone Joint Res. 2018 Aug 4;7(7):476-484
pubmed: 30123497
JBJS Case Connect. 2021 Mar 8;11(1):
pubmed: 33684083
J Am Acad Orthop Surg. 2020 Sep 15;28(18):e803-e809
pubmed: 32520902
Spine (Phila Pa 1976). 2020 Jul 1;45(13):872-876
pubmed: 32539289
Acta Orthop. 2021 Feb;92(1):120-127
pubmed: 33106069
Acta Orthop. 2019 Feb;90(1):81-87
pubmed: 30371122
Spine Surg Relat Res. 2019 Oct 20;4(2):148-151
pubmed: 32405561
BMC Musculoskelet Disord. 2020 Aug 5;21(1):519
pubmed: 32758204
Acta Orthop. 2021 Aug;92(4):479-484
pubmed: 33757381
J Pediatr Orthop B. 2020 Nov;29(6):611-617
pubmed: 31904740
Biomed Res Int. 2017;2017:8032510
pubmed: 29130046
J Clin Orthop Trauma. 2020 Jun 24;14:151-155
pubmed: 33717907
Int Orthop. 2017 Sep;41(9):1915-1923
pubmed: 28389837
Bone Joint Res. 2021 Jul;10(7):425-436
pubmed: 34269599
Spine (Phila Pa 1976). 2020 Feb 1;45(3):170-176
pubmed: 31513114
Spine Deform. 2020 Aug;8(4):763-770
pubmed: 32170659
Med Devices (Auckl). 2020 Mar 17;13:75-85
pubmed: 32256128
Int Orthop. 2020 Sep;44(9):1773-1783
pubmed: 32494843