Completely epicardial implantable cardioverter/defibrillator (ICD) and CRT-D systems: A case series and systematic literature review.


Journal

Pacing and clinical electrophysiology : PACE
ISSN: 1540-8159
Titre abrégé: Pacing Clin Electrophysiol
Pays: United States
ID NLM: 7803944

Informations de publication

Date de publication:
Sep 2021
Historique:
revised: 08 06 2021
received: 09 04 2021
accepted: 11 07 2021
pubmed: 22 7 2021
medline: 1 2 2022
entrez: 21 7 2021
Statut: ppublish

Résumé

Epicardial ICD systems and CRT-Ds using high voltage coils represent an alternative to transvenous systems in patients without central venous access and prior device complications including infection. We present a case series in the adult population of epicardial ICD/CRTD systems using high voltage epicardial coils. We summarize the existing data regarding techniques, efficacy, and safety. A retrospective board approved medical record review was conducted for all patients undergoing epicardial ICD/CRTD placement at our institution between January 2010 and May 2020. The literature was reviewed for prior published trials, case reports, and case series of epicardial high voltage coil insertions. Eleven patients (six female, mean age 48 years) underwent epicardial ICD/CRTD implant including 5/11 completely epicardial CRTD systems. The procedure was performed via median sternotomy in eight patients, left anterior thoracotomy in two patients, and sub-xiphoid approach in one patient. After a mean follow up of 35 months, appropriate successful shocks were delivered in two (18%) patients and no patients received an inappropriate shock. Three of five (60%) patients had volumetric remodeling with CRT with significant improvement of LV EF. Device-related complications requiring a surgical/percutaneous revision or another DFT test occurred in six patients (54%). One patient died during follow up due to refractory heart failure. No cases of epicardial device infection, coronary artery compression, constrictive pericarditis, or erosion of defibrillator coils into intrathoracic organs were reported. No randomized studies comparing safety and efficacy of traditional transvenous or subcutaneous ICD systems and epicardial ICD systems using contemporary high voltage coils were found nor any studies directly comparing epicardial defibrillator patches versus epicardial coils. Thirteen case series and 24 single case reports published between 2004 and 2020 were identified describing in total a heterogenous group of 188 patients with ICD systems incorporating one or more epi- or pericardially positioned shock coils. The use of epicardial defibrillation coils for ICD/CRT-D is a feasible treatment option for patients with either failed or contraindicated transvenous ICD systems. Dedicated epicardial high voltage leads with integrated pace-sense electrodes and specialized delivery tools for minimal invasive implantations may improve longer term outcomes.

Sections du résumé

BACKGROUND BACKGROUND
Epicardial ICD systems and CRT-Ds using high voltage coils represent an alternative to transvenous systems in patients without central venous access and prior device complications including infection.
OBJECTIVE OBJECTIVE
We present a case series in the adult population of epicardial ICD/CRTD systems using high voltage epicardial coils. We summarize the existing data regarding techniques, efficacy, and safety.
METHODS METHODS
A retrospective board approved medical record review was conducted for all patients undergoing epicardial ICD/CRTD placement at our institution between January 2010 and May 2020. The literature was reviewed for prior published trials, case reports, and case series of epicardial high voltage coil insertions.
RESULTS RESULTS
Eleven patients (six female, mean age 48 years) underwent epicardial ICD/CRTD implant including 5/11 completely epicardial CRTD systems. The procedure was performed via median sternotomy in eight patients, left anterior thoracotomy in two patients, and sub-xiphoid approach in one patient. After a mean follow up of 35 months, appropriate successful shocks were delivered in two (18%) patients and no patients received an inappropriate shock. Three of five (60%) patients had volumetric remodeling with CRT with significant improvement of LV EF. Device-related complications requiring a surgical/percutaneous revision or another DFT test occurred in six patients (54%). One patient died during follow up due to refractory heart failure. No cases of epicardial device infection, coronary artery compression, constrictive pericarditis, or erosion of defibrillator coils into intrathoracic organs were reported. No randomized studies comparing safety and efficacy of traditional transvenous or subcutaneous ICD systems and epicardial ICD systems using contemporary high voltage coils were found nor any studies directly comparing epicardial defibrillator patches versus epicardial coils. Thirteen case series and 24 single case reports published between 2004 and 2020 were identified describing in total a heterogenous group of 188 patients with ICD systems incorporating one or more epi- or pericardially positioned shock coils.
CONCLUSION CONCLUSIONS
The use of epicardial defibrillation coils for ICD/CRT-D is a feasible treatment option for patients with either failed or contraindicated transvenous ICD systems. Dedicated epicardial high voltage leads with integrated pace-sense electrodes and specialized delivery tools for minimal invasive implantations may improve longer term outcomes.

Identifiants

pubmed: 34287989
doi: 10.1111/pace.14318
doi:

Types de publication

Case Reports Journal Article Systematic Review

Langues

eng

Sous-ensembles de citation

IM

Pagination

1616-1630

Informations de copyright

© 2021 The Authors. Pacing and Clinical Electrophysiology published by Wiley Periodicals LLC.

Références

Ozyuksel A, Ersoy C, Akdeniz C, et al. Implantation of looped epicardial cardioverter defibrillator coil on the surface of the right ventricular outflow tract. J Card Surg. 2015 Apr; 30:376-380.
Snyder CS, Lucas V, Young T, et al. Minimally invasive implantation of a cardioverter-defibrillator in a small patient. J Thorac Cardiovasc Surgery. 2007;133:1375-1376.
Bhakta M, Obioha CC, Sorajja D, et al. Non-traditional ICD placement in adult patients with limited venous access: a case series. Pacing Clin Electrophysiol. 2010;33:217-225.
Stephenson EA, Batra AS, Knilans TK, et al. A multicentre experience with novel ICD configurations in the paediatric and congenital heart disease population. J Cardiovasc Electrophysiol. 2006;17:41-46.
Haydin S, Saygi M, Ergul Y. Subxiphoid approach to epicardial implantation of implantable cardioverter defibrillators in children. Pacing Clin Electrophysiol. 2013 Aug; 36:926-930.
Molina JE, Benditt DG. Epicardial subxiphoid implantable defibrillator lead: superior effectiveness after failure of standard implants. PACE. 2004;27:1500-1506.
Sony J, Liebermann RA, et al. Percutaneous epicardial defibrillation coil implantation: a viable technique to manage refractory defibrillation threshold. Images and Case reports in Arrhythmia and Electrophysiology. Circ Arrhythm Electrophysiol. 2010;3:214-217.
Curnis A, Muneretto C, Bisleri G, et al. Thoracoscopic Implantation of an Array Electrode in the Pericardium Transverse Sinus to Reduce Defibrillation Threshold. Innovations. 2017 Aug; 12:e6-e9.
Ip JE, Krishnan U, Girardi LN, Lerman BB, A hybrid endocardial-epicardial biventricular implantable cardioverter defibrillator to circumvent the tricuspid valve. PACE 2020 Oct; Online ahead of print.
Jaroszewski DE, Altemose GT, Scott LR. Non-traditional surgical approaches for implantation of pacemaker and cardioverter defibrillator systems in patients with limited venous access. Ann Thorac Surg. 2009 Jul; 88:112-116.
Shalaby A, Sharma M, Zenati M. Robotic Implantation of a Multichamber Cardiac Resychronization Therapy Defibrillator. PACE. 2006:906-909.
Amraoui S, Labrousse L, Jansens J, et al. Implantation of a cardiac resynchronization therapy and defibrillation system using a robotic approach. Heart Rhythm Case reports. 2015;1:356-359.
Sugumar H, Matalanis G, Toner L, O'Donnell D. Epicardial CRT-D in a Pacing Dependant Patient Without Transvenous Options. Heart Lung Circ. 2017;26:e48-e49.
Behar JM, Islam S, Adams B, et al. A completely epicardial biventricular defibrillator for a pacing dependent patient with no superior central venous access. JACC. 2018;4:277-279.
Wey HE, Chua K, Balkhy H, et al. Physiological optimization of robotic endoscopic epicardial CRT-D Implantation using multielectrode electroanatomic mapping. J Cardiovasc Electrophysiol. 2019;30:2564-2568.
Starck C, Mayer K, Hürlimann D, et al. Completely Epicardial Implantation of a Cardiac Resynchronization Therapy Defibrillator Using a Minimal Invasive Approach. Thorac Cardiovasc Surg. 2014;62:70-72.
Sughimoto K, Tsuchida Y, Hayashi H, et al. Epicardial implantable Cardioverter-Defibrillator in a 2-month-old infant. Ann Thorac Surg. 2017;103:e263-5.
Amraoui S, Ploux S, Sacher F, et al. Epicardial Defibrillation System Implantation: long term Follow-up. Heart Rhythm. 2018;15. Supplement: S12 (Abstract Heart rhythm congress B-AB05-03).
Mah DY, Prakash A, Porras D, et al. Coronary artery compression from epicardial leads: more common than we think. Heart rhythm. 2018;15:1439-1447.
Amraoui S, Ploux S, Sacher F, et al. Epicardial Defibrillation System Implantation: long term Follow-up. Heart Rhythm. 2018;15. Supplement: S12 (Abstract Heart rhythm congress B-AB05-03.
Haydin S, Saygi M, Ergul Y. Subxiphoid approach to epicardial implantation of implantable cardioverter defibrillators in children. Pacing Clin Electrophysiol. 2013;36:926-930.
Zipes DP, Roberts D. Results of the international study of the implantable pacemaker cardioverter-defibrillator. A comparison of epicardial and endocardial lead systems. The Pacemaker-Cardioverter-Defibrillator Investigators. Circulation. 1995;92:59-65.
Korte T, Jung W, Spehl S, et al. Incidence of ICD lead related complications during long term follow up: comparison of epicardial and endocardial electrode systems. Pacing Clin Electrophysiol. 1995;18:2053-2061.
Tjong FV, Brouwer TF, Smeding L, et al. Combined leadless pacemaker and subcutaneous implantable defibrillator therapy: feasibility, safety and performance. Europace. 2016;18:1740-1747.
Erath JW, Sirat AS, Vamos M, Hohnloser SH. Epicardial CRT-P and S-ICD Implantation in a Young Patient with Persistent Left superior Vena Cava. Herzschrittmacherther Electrophysiol. 2016;27:396-398.
Sidhu BS, Gould J, Porter B, et al. Completely Leadless Cardiac Resynchronization Defibrillator System. JACC: Clin Electrophy. 2020;6:588-589.
Higgins SL. Biventricular ICD placement percutaneously via the Iliac Vein: case Reports and Review. J Innov Card Rhythm Manag. 2017;8:2784-2789.
Myung-Jin C, Jae-Sun U, Tae Hoon K, et al. Two cases of transhepatic Implantation of Cardia Implantable Electronic Device: all Roads lead to Rome. Int J Arrhythm. 2017;18:209-214.
Elayi CS, Allen CL, Leung S, et al. Inside-out access: a new method of lead placement for patients with central venous occlusions. Heart Rhythm. 2011;8:851e7.
Cannon BC, Friedman RA, Fenrich AL, et al. Innovative Techniques for Placement of Implantable Cardioverter-Defibrillator Leads in Patients with limited venous access to the heart. Pacing and clinical electrophysiology. 2006;29:181-187.
Molina JE. Surgical options for endocardial lead placement when upper veins are obstructed or non usable. J Interv Card Electrophysiol. 2004;11:149-154.
Rodriguez-Manero M, Kreidieh B, Ibarra-Cortez S, et al. Coronary vein defibrillator coil placement in patients with high defibrillation thresholds. J Arrhythm. 2019;35:79-85.
Cooper JA, Latacha MP, Soto GE, et al. The azygos defibrillator lead for elevated defibrillation thresholds: implant technique, lead stability and patient series. Pacing Clin Electrophysiol. 2008;31:1405-1410.
Cesario D, Bhargava M, Valderrabano M, et al. Azygos vein lead implantation: a novel adjunctive technique for implantable cardioverter defibrillator placement. J Cardiovasc Electrophysiol. 2004;15:780-783.
Nilsson KR, Jackson KP. Hemiazygous coil placement for high-defibrillation thresholds in a patient with a right-sided implantable cardioverter defibrillator. Pacing Clin Electrophysiol. 2012;35:e10-2.
Markewitz A, Kaulbach H, Mattke S, et al. Influence of anodal electrode position on transvenous defibrillation efficacy in humans: a prospective randomized comparison. Pacing Clin Electrophysiol. 1997;20:2193-2199. Pt 1.
Brady PA, Friedmann PA, Trusty JM, et al. High failure rate for an epicardial ICD lead: implications for long term follow up of patients with an ICD. J Am Coll Cardiol. 1998;31:616-622.
Mah DY, Prakash A, Porras D, et al. Coronary artery compression from epicardial leads: more common than we think. Heart rhythm. 2018;15:1439-1447.
Carreras EM, Duncan WJ, Djurdjev O, et al. Cardia strangulation following epicardial pacemaker implantation: a rare pediatric complication. J Thorac Cardiovasc Surg. 2015;149:522-527.
Takeuchi D, Tomizawa Y. Cardiac Strangulation from epicardial pacemaker leads: diagnosis, treatment and prevention. Gen Thorac Cardiovasc Surg. 2015;63:22-29.
Chevalier P, Moncada E, Canu G, et al. Symptomatic pericardial disease associated with patch electrodes of the automatic implantable cardioverter defibrillator: an underestimated complication?. Pacing Clin Electrophysiol. 1996;19:2150-2152. Pt 1.
Mendonca Costa C, Neic A, Kerfoot E, et al. Pacing in proximity to scar during cardiac resynchronization therapy increases local dispersion of repolarization and susceptibility to ventricular arrhythmogenesis. Heart Rhythm. 2019;16:1475-1483.
Nathan S, Piccione W, Kumar A, et al. Erosion of an Extrapericardial Implantable Cardioverter Defibrillator Patch Through the Gastric Fundus with Fistulous Tract Formation. Cardiology in Review. 2006;14:e21-e23.
Nolan RL, McAdams HP. Bronchopericardial fistula after placement of an automatic implantable cardioverter defibrillator: radiographic and CT findings. Am J Roentgenology. 1999;172:365-368.
Vuorinen AM, Pakarinen S, Jaakkola I, et al. Clinical experience of magnetic resonance imaging in patients with cardiac pacing devices: unrestricted patient population. Acta Radiol. 2019;19.
Shah AD, et al. MRI safety in nonconditional pacemaker and defibrillator recipients: a meta analysis and systematic review. Heart Rhythm. 2018;15:1001-1008.
Balmer C, Gass M, Dave H, et al. Magnetic resonance imaging of patients with epicardial leads: in vitro evaluation of temperature changes at the lead tip. J Int Car Electrophysiology. 2019;56:321.326.
Rickard J, Johnston DR, Price J, et al. Reverse ventricular remodelling and long term survival in patients undergoing cardiac resynchronization with surgically versus percutaneously placed left ventricular pacing leads. Heart Rhythm. 2015;12:517-523.
Ailawadi G, LaPar DJ, Swenson BR, et al. Surgically placed left ventricular leads provide similar outcomes to percutaneous leads in patients with failed coronary sinus lead placement. Heart Rhythm. 2010;7:619-625.

Auteurs

Johanna B Tonko (JB)

Cardiothoracic Unit, Guy's and St.Thomas' NHS Foundation Trust, London, UK.
Faculty of Life Sciences and Medicine, King's College London, London, UK.

Christopher Blauth (C)

Cardiothoracic Unit, Guy's and St.Thomas' NHS Foundation Trust, London, UK.

Eric Rosenthal (E)

Faculty of Life Sciences and Medicine, King's College London, London, UK.
Evelina Children's Hospital, Department of Paediatric Cardiology, St Thomas Hospital, London, UK.

Christopher A Rinaldi (CA)

Cardiothoracic Unit, Guy's and St.Thomas' NHS Foundation Trust, London, UK.
Faculty of Life Sciences and Medicine, King's College London, London, UK.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH