Assessment of predictive indicators of acute gastrointestinal toxicity using in vivo transmission images.
in vivo EPID
outcomes analysis
toxicity
Journal
Medical physics
ISSN: 2473-4209
Titre abrégé: Med Phys
Pays: United States
ID NLM: 0425746
Informations de publication
Date de publication:
Dec 2021
Dec 2021
Historique:
revised:
08
09
2021
received:
29
03
2021
accepted:
09
10
2021
pubmed:
20
10
2021
medline:
17
12
2021
entrez:
19
10
2021
Statut:
ppublish
Résumé
For pelvic and abdominal treatments, excess dose to the bowel can result in acute toxicities. Current estimates of bowel toxicity are based on pre-treatment dose-volume histogram data. However, the actual dose the bowel receives depends on interfraction variations, such as patient anatomy changes. We propose a method to model bowel toxicities, incorporating in vivo patient information using transit electronic portal imaging device (EPID) images. For 63 patients treated to the lower thorax, abdomen, or pelvis on the Varian Halcyon, weekly chart review was performed to obtain incidences of grade 2 or higher toxicity, RTOG scale. Twenty patients presented with acute gastrointestinal (GI) toxicity. All patients were treated with conventional fractionation. For each treatment plan, the absolute volume dose-volume histogram of the bowel was exported and analyzed. Additionally, for each fraction of treatment, in vivo EPID images were collected and used to estimate the change in radiation transmission during the course of treatment. A logistic model was used to test correlations between acute GI toxicity and bowel dosimetric parameters as well as metrics obtained from in vivo image measurements. After performing the fit to the in vivo EPID data, the bootstrap resampling method was used to create confidence intervals. In vivo EPID image metrics from an additional 42 patients treated to the lower thorax, abdomen, or pelvis were used to validate the logistic model fit. The incidence of toxicity versus the volume of 40 Gy to the bowel space was fitted with a logistic function, which was superior to an average model (p < 0.0001) and agrees with previously published models. For the initial in vivo EPID data, the incidence of toxicity versus the sum of in vivo transmission measurements showed marginal significance after 15 fractions (p = 0.10) of treatment and a significance of p = 0.038 is seen at the 20th fraction, when compared to an average model. For the validation data set, the logistic model of the in vivo transmission measurement after 20 fractions was superior to the average model (p = 0.043), with the model falling within the 68% confidence interval of the fit of the initial data set. Dose-volume constraints to reduce the incidence of acute GI toxicity have been validated. The presented novel EPID transmission-based metric can be used to identify GI toxicity as patients progress through treatment.
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
8152-8162Informations de copyright
© 2021 American Association of Physicists in Medicine.
Références
Minsky BD, Conti JA, Huang Y, Knopf K. Relationship of acute gastrointestinal toxicity and the volume of irradiated small bowel in patients receiving combined modality therapy for rectal cancer. J Clin Oncol. 1995;13(6):1409-1416.
Devisetty K, Mell LK, Salama JK, et al. A multi-institutional acute gastrointestinal toxicity analysis of anal cancer patients treated with concurrent intensity-modulated radiation therapy (IMRT) and chemotherapy. Radiother Oncol. 2009;93(2):298-301.
Tho LM, Glegg M, Paterson J, et al. Acute small bowel toxicity and preoperative chemoradiotherapy for rectal cancer: investigating dose-volume relationships and role for inverse planning. Int J Radiat Oncol Biol Phys. 2006;66(2):505-513.
Robertson JM, Lockman D, Yan D, Wallace M. The dose-volume relationship of small bowel irradiation and acute grade 3 diarrhea during chemoradiotherapy for rectal cancer. Int J Radiat Oncol Biol Phys. 2008;70(2):413-418.
Perna L, Alongi F, Fiorino C, et al. Predictors of acute bowel toxicity in patients treated with IMRT whole pelvis irradiation after prostatectomy. Radiother Oncol. 2010;97(1):71-75.
Simpson DR, Song WY, Moiseenko V, et al. Normal tissue complication probability analysis of acute gastrointestinal toxicity in cervical cancer patients undergoing intensity modulated radiation therapy and concurrent cisplatin. Int J Radiat Oncol Biol Phys. 2012;83(1):e81-e86.
Roeske JC, Bonta D, Mell LK, Lujan AE, Mundt AJ. A dosimetric analysis of acute gastrointestinal toxicity in women receiving intensity-modulated whole-pelvic radiation therapy. Radiother Oncol. 2003;69(2):201-207.
Zhen H, Nelms BE, Tomé WA. On the use of biomathematical models in patient-specific IMRT dose QA. Med Phys. 2013;40(7):071702.
Sumida I, Yamaguchi H, Kizaki H, et al. Novel radiobiological gamma index for evaluation of 3-dimensional predicted dose distribution. Int J Radiat Oncol Biol Phys. 2015;92(4):779-786.
Shelley LEA, Scaife JE, Romanchikova M, et al. Delivered dose can be a better predictor of rectal toxicity than planned dose in prostate radiotherapy. Radiother Oncol. 2017;123(3):466-471.
Burnet NG, Scaife JE, Romanchikova M, et al. Applying physical science techniques and CERN technology to an unsolved problem in radiation treatment for cancer: the multidisciplinary ‘VoxTox’ research programme. Cern ideaSquare J Exp Innov. 2017;1(1):3-12.
Alam S, Thor M, Rimner A, et al. Quantification of accumulated dose and associated anatomical changes of esophagus using weekly magnetic resonance imaging acquired during radiotherapy of locally advanced lung cancer. Phys Imaging Radiat Oncol. 2020;13:36-43.
Casares-Magaz O, Bülow S, Pettersson NJ, et al. High accumulated doses to the inferior rectum are associated with late gastro-intestinal toxicity in a case-control study of prostate cancer patients treated with radiotherapy. Acta Oncologica. 2019;58(10):1543-1546.
Shelley LEA, Sutcliffe MPF, Thomas SJ, et al. Associations between voxel-level accumulated dose and rectal toxicity in prostate radiotherapy. Phys Imaging Radiat Oncol. 2020;14:87-94.
Shelley LE. Accumulating delivered dose to the rectum to improve toxicity prediction in prostate radiotherapy. Doctoral thesis. University of Cambridge; 2019.
Anders M, Mijnheer BJ. Clinical 3D Dosimetry in Modern Radiation Therapy. 1st ed. CRC Press; 2018.
Mans A, Wendling M, McDermott LN, et al. Catching errors with in vivo EPID dosimetry. Med Phys. 2010;37(6):2638-2644.
Olaciregui-Ruiz I, Rozendaal R, Mijnheer B, Mans A. Site-specific alert criteria to detect patient-related errors with 3D EPID transit dosimetry. Med Phys. 2019;46(1):45-55.
Persoon LCGG, Nijsten SMJJG, Wilbrink FJ, et al. Interfractional trend analysis of dose differences based on 2D transit portal dosimetry. Phys Med Biol. 2012;57(20):6445-6458.
Nijsten SMJJG, van Elmpt WJC, Mijnheer BJ, et al. Prediction of DVH parameter changes due to setup errors for breast cancer treatment based on 2D portal dosimetry. Med Phys. 2009;36(1):83-94.
Rozendaal RA, Mijnheer BJ, Van Herk M, Mans A. In vivo portal dosimetry for head-and-neck VMAT and lung IMRT: linking γ-analysis with differences in dose-volume histograms of the PTV. Radiother Oncol. 2014;112(3):396-401.
Kavanagh BD, Pan CC, Dawson LA, et al. Radiation dose-volume effects in the stomach and small bowel. Int J Radiat Oncol Biol Phys. 2010;76(3):S101-S107.
Radiation Therapy Oncology Group. Acute radiation morbidity scoring criteria. National Cancer Instutute; 2008.
Owen WJ, Cowan G. Statistical data analysis. Technometrics. 2000;42(3):311-312.
Steers JM, Zavala Bojorquez J, Moore K, Bojechko C. Using in vivo EPID images to detect and quantify patient anatomy changes with gradient dose segmented analysis. Med Phys. 2020;47(11):5419-5427.
Moiseenko V, Marks LB, Grimm J, et al. A primer on dose-response data modeling in radiation therapy. Int J Radiat Oncol Biol Phys. 2021;110(1):11-20.
Scott AJ, Hosmer DW, Lemeshow S. Applied logistic regression. Biometrics. 1991;47(4):1632-1633.
Bentzen SM. Radiobiological considerations in the design of clinical trials. Radiother Oncol. 1994;32(1):1-11.
Jaffray DA, Lindsay PE, Brock KK, Deasy JO, Tomé WA. Accurate accumulation of dose for improved understanding of radiation effects in normal tissue. Int J Radiat Oncol Biol Phys. 2010;76(3):S135-S139.
Olaciregui-Ruiz I, Rozendaal R, Mijnheer B, Van Herk M, Mans A. Automatic in vivo portal dosimetry of all treatments. Phys Med Biol. 2013;58(22):8253-8264.