Caution: Air Embolism Related to Heli-Fx EndoAnchor System in Zone 2 Thoracic Aneurysm Repair.
Aorta, Thoracic
Aortic Aneurysm, Abdominal
/ surgery
Aortic Aneurysm, Thoracic
/ complications
Blood Vessel Prosthesis
/ adverse effects
Blood Vessel Prosthesis Implantation
/ adverse effects
Embolism, Air
/ diagnostic imaging
Endoleak
/ etiology
Endovascular Procedures
/ adverse effects
Humans
Prosthesis Design
Risk Factors
Treatment Outcome
EndoAnchors
Heli-Fx
air embolism
complications
endoleak
thoracic endovascular aortic repair
Journal
Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists
ISSN: 1545-1550
Titre abrégé: J Endovasc Ther
Pays: United States
ID NLM: 100896915
Informations de publication
Date de publication:
10 2022
10 2022
Historique:
pubmed:
3
12
2021
medline:
9
9
2022
entrez:
2
12
2021
Statut:
ppublish
Résumé
We report a case of significant air embolization to the ascending aorta immediately following deployment of EndoAnchors in the aortic arch during a procedure to correct a type 1A endoleak. The novel Heli-Fx EndoAnchor system (Medtronic Vascular, Santa Rosa, CA, USA) was used to deploy helical anchors in the distal aortic arch during a procedure to correct a type 1A endoleak following Zone 2 thoracic endovascular aortic repair of a saccular proximal descending thoracic aorta aneurysm (DTAA). The patient developed ST-segment elevations principally in the inferior leads and severe hypotension moments after EndoAnchor deployment at the proximal edge of the endograft. Transesophageal echocardiogram revealed severe right ventricular hypokinesis and a large amount of air in the ascending aorta. Subsequent management and clinical and radiological 30-day follow-up is presented in addition to a review of the literature and ex vivo testing with the Heli-Fx system to examine potential causes and solutions. Precautions, such as pressurized saline infusion to the side port of guiding sheath, should be used whenever manipulating catheters and sheaths such as the EndoAnchor system in the aortic arch to prevent this potentially lethal complication.