Local diagnostic reference levels (LDRLs) for full-field digital mammography (FFDM) and digital breast tomosynthesis (DBT) procedures in Morocco.
Journal
Journal of medical imaging and radiation sciences
ISSN: 1876-7982
Titre abrégé: J Med Imaging Radiat Sci
Pays: United States
ID NLM: 101469694
Informations de publication
Date de publication:
06 2022
06 2022
Historique:
received:
13
01
2022
accepted:
10
03
2022
pubmed:
3
4
2022
medline:
25
5
2022
entrez:
2
4
2022
Statut:
ppublish
Résumé
The objective of this study was to establish local diagnostic reference levels (LDRLs) for the full-field digital mammography (FFDM) and tomosynthesis (DBT) in Moroccan health facilities. Data from 146 women were collected from three facilities. The proposed DRLs were defined as the 75th percentile of the mean average glandular dose (AGD) distribution. The mean AGD recorded in this study for the three centers was 1.47 mGy for all centers, and 1.42 mGy and 1.64 mGy for the CC and MLO projections, respectively. The mean compressed breast thickness (CBT) values recorded in this current study were 55 mm, the LDRLs reported for all centers was 1.7 mGy, the CC projection was 1.6 mGy, and the MLO projection was 1.8 mGy. In addition, the LDRLs reported in the current study were compared with those from previous studies for other countries, including the United Kingdom, Japan, Ghana, and Sri Lanka. This work provides an assessment of local DRLs for mammography in Morocco and is suggested as a starting point that will allow professionals to evaluate and optimize their practice. Furthermore, the definition of national DRLs is a necessary process in optimizing Moroccan medical exposures.
Identifiants
pubmed: 35365437
pii: S1939-8654(22)00049-2
doi: 10.1016/j.jmir.2022.03.008
pii:
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
242-247Informations de copyright
Copyright © 2022. Published by Elsevier Inc.