Migration of Subtalar Implants in Paediatric Flatfoot Correction: A Pilot Study in Synthetic Bone Models.

minimally invasive surgery paediatric flatfoot subtalar arthroereisis subtalar implant

Journal

Ortopedia, traumatologia, rehabilitacja
ISSN: 2084-4336
Titre abrégé: Ortop Traumatol Rehabil
Pays: Poland
ID NLM: 101240146

Informations de publication

Date de publication:
30 Apr 2022
Historique:
entrez: 13 5 2022
pubmed: 14 5 2022
medline: 18 5 2022
Statut: ppublish

Résumé

Subtalar implant migration as a complication following subtalar arthroeresis has been described in the scientific literature. However, clinical studies do not allow for unequivocally determining the underlying causes. The aim of the study is to determine the risk of migration of two geometric types of subtalar implants. Biomechanical tests were carried out on a synthetic bone model with a soft tissue substitute, which allowed for reduction of variability of results caused by biological differences. A foot model mirroring natural anatomy was made from synthetic bone and a silicone soft tissue substitute with the same hardness as that of the soft tissues of the foot. Two types of 11 mm Ti6Al4V titanium alloy implants were studied, namely, a rectangular subtalar screw and a cylindrical subtalar screw, a type commonly used in flatfoot reconstruction surgery in children. The screws were placed in the sinus tarsi and subjected to cyclic loading (up to 1,000,000 cycles at a frequency of 5 Hz, with a maximum load of 500 N). Comparative pull-out force tests were performed immediately following implantation and after the dynamic loading test. Wyniki. Following the dynamic loading test, all 12 samples were qualified for the pull-out force test. Cylindrical screws demonstrated higher pull-out force values both for the samples tested immediately following implantation and for those that underwent dynamic loading. Implants of the same shape did not show statistically significant differences in the Mann-Whitney U test (p >0.05). Wniosek. The synthetic research model produces reproducible results in the assessment of risk of implant migration. Long-term loading does not significantly affect the risk of implant migration.

Sections du résumé

BACKGROUND BACKGROUND
Subtalar implant migration as a complication following subtalar arthroeresis has been described in the scientific literature. However, clinical studies do not allow for unequivocally determining the underlying causes. The aim of the study is to determine the risk of migration of two geometric types of subtalar implants. Biomechanical tests were carried out on a synthetic bone model with a soft tissue substitute, which allowed for reduction of variability of results caused by biological differences.
MATERIAL AND METHODS METHODS
A foot model mirroring natural anatomy was made from synthetic bone and a silicone soft tissue substitute with the same hardness as that of the soft tissues of the foot. Two types of 11 mm Ti6Al4V titanium alloy implants were studied, namely, a rectangular subtalar screw and a cylindrical subtalar screw, a type commonly used in flatfoot reconstruction surgery in children. The screws were placed in the sinus tarsi and subjected to cyclic loading (up to 1,000,000 cycles at a frequency of 5 Hz, with a maximum load of 500 N). Comparative pull-out force tests were performed immediately following implantation and after the dynamic loading test. Wyniki. Following the dynamic loading test, all 12 samples were qualified for the pull-out force test. Cylindrical screws demonstrated higher pull-out force values both for the samples tested immediately following implantation and for those that underwent dynamic loading. Implants of the same shape did not show statistically significant differences in the Mann-Whitney U test (p >0.05). Wniosek. The synthetic research model produces reproducible results in the assessment of risk of implant migration. Long-term loading does not significantly affect the risk of implant migration.

Identifiants

pubmed: 35550354
doi: 10.5604/01.3001.0015.8374
pii: 01.3001.0015.8374
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

121-132

Auteurs

Michał Kwiatkowski (M)

Krajowe Międzysektorowe Studia Doktoranckie na Uniwersytecie Medycznym w Białymstoku, Polska / National Intersectoral Doctoral Studies at the Medical University of Białystok, Poland.

Tomasz Guszczyn (T)

Klinika Ortopedii i Traumatologii Dziecięcej, Uniwersytet Medyczny w Białymstoku, Polska / Department of Pediatric Orthopaedics and Traumatology, Medical University of Białystok, Poland.

Andrzej Sobolewski (A)

ChM Sp. z o.o., Polska / ChM Sp. z o.o., Poland.

Marcin Sar (M)

Oddział urazowo-ortopedyczny, SP ZOZ Bielsk Podlaski, Polska / Department of Orthopaedics and Traumatology, SP ZOZ Bielsk Podlaski, Poland.

Adam Hermanowicz (A)

Klinika Chirurgii i Urologii Dziecięcej, Uniwersytet Medyczny w Białymstoku, Polska / Department of Pediatric Surgery and Urology, Medical University of Białystok, Poland.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH