A Real-Time Energy Monitoring System for an MRI Hybrid Ablation System.
Journal
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
ISSN: 2694-0604
Titre abrégé: Annu Int Conf IEEE Eng Med Biol Soc
Pays: United States
ID NLM: 101763872
Informations de publication
Date de publication:
07 2022
07 2022
Historique:
entrez:
10
9
2022
pubmed:
11
9
2022
medline:
14
9
2022
Statut:
ppublish
Résumé
The MRI hybrid ablation system is an approach to use the MR (magnetic resonance) scanner's radiofrequency amplifier itself as power source for ablation. Hereby, an electrode is connected to the MR internal radiofrequency amplifier. An average RF power is provided through a train of short RF pulses, which is sufficient to thermally destroy tissue. However, ablation with too high power values can cause tissue carbonizations, thus impeding the ablation procedure. Therefore, monitoring of the energy and the power absorbed inside the tissue is necessary. For this purpose, a measurement system was designed to measure the energy applied to the tissue when using the concept of an MRI hybrid ablation system. The system consists of a dual-directional coupler, RF-to-RMS sensors, and a microcontroller. The gradient calculation of the measured energy curve provides information about the absorbed RF power in the tissue. Validation measurements of the system were performed and compared with measurements from the MR-internal power measurement system. The energy monitoring system was also tested in an ablation experiment with ex-vivo animal tissue using the MRI hybrid ablation system. The measurements showed that the applied RF power can be monitored in real-time. It has been shown that the mean RF power absorbed in the patient decreased during an ablation procedure due to an occurring impedance mismatch and tissue changes. In further work, the influence of the monitoring system on the quality of the MR images should be investigated. Clinical relevance- This paper demonstrates an energy monitoring system for an RF ablation system, which can be used inside an MR environment.
Identifiants
pubmed: 36086647
doi: 10.1109/EMBC48229.2022.9870997
doi:
Types de publication
Journal Article
Research Support, Non-U.S. Gov't
Langues
eng
Sous-ensembles de citation
IM