Simplified modeling of implanted medical devices with metallic filamentary closed loops exposed to low or medium frequency magnetic fields.

In silico modeling Magnetic hyperthermia Magnetic resonance imaging Metallic implants Skull grid Stent Transcranial magnetic stimulation

Journal

Computer methods and programs in biomedicine
ISSN: 1872-7565
Titre abrégé: Comput Methods Programs Biomed
Pays: Ireland
ID NLM: 8506513

Informations de publication

Date de publication:
Feb 2023
Historique:
received: 28 06 2022
revised: 12 12 2022
accepted: 16 12 2022
pubmed: 26 12 2022
medline: 8 2 2023
entrez: 25 12 2022
Statut: ppublish

Résumé

Electric currents are induced in implanted medical devices with metallic filamentary closed loops (e.g., fixation grids, stents) when exposed to time varying magnetic fields, as those generated during certain diagnostic and therapeutic biomedical treatments. A simplified methodology to efficiently compute these currents, to estimate the altered electromagnetic field distribution in the biological tissues and to assess the consequent biological effects is proposed for low or medium frequency fields. The proposed methodology is based on decoupling the handling of the filamentary wire and the anatomical body. To do this, a circuital solution is adopted to study the metallic filamentary implant and this solution is inserted in the electromagnetic field solution involving the biological tissues. The Joule losses computed in the implant are then used as a forcing term for the thermal problem defined by the bioheat Pennes' equation. The methodology is validated against a model problem, where a reference solution is available. The proposed simplified methodology is proved to be in good agreement with solutions provided by alternative approaches. In particular, errors in the amplitude of the currents induced in the wires result to be always lower than 3%. After the validation, the methodology is applied to check the interactions between the magnetic field generated by different biomedical devices and a skull grid, which represents a complex filamentary wire implant. The proposed simplified methodology, suitable to be applied to closed loop wires in the low to intermediate frequency range, is found to be sufficiently accurate and easy to apply in realistic exposure scenarios. This modeling tool allows analyzing different types of small implants, from coronary and biliary duct stents to orthopedic grids, under a variety of exposure scenarios.

Sections du résumé

BACKGROUND AND OBJECTIVES OBJECTIVE
Electric currents are induced in implanted medical devices with metallic filamentary closed loops (e.g., fixation grids, stents) when exposed to time varying magnetic fields, as those generated during certain diagnostic and therapeutic biomedical treatments. A simplified methodology to efficiently compute these currents, to estimate the altered electromagnetic field distribution in the biological tissues and to assess the consequent biological effects is proposed for low or medium frequency fields.
METHODS METHODS
The proposed methodology is based on decoupling the handling of the filamentary wire and the anatomical body. To do this, a circuital solution is adopted to study the metallic filamentary implant and this solution is inserted in the electromagnetic field solution involving the biological tissues. The Joule losses computed in the implant are then used as a forcing term for the thermal problem defined by the bioheat Pennes' equation. The methodology is validated against a model problem, where a reference solution is available.
RESULTS RESULTS
The proposed simplified methodology is proved to be in good agreement with solutions provided by alternative approaches. In particular, errors in the amplitude of the currents induced in the wires result to be always lower than 3%. After the validation, the methodology is applied to check the interactions between the magnetic field generated by different biomedical devices and a skull grid, which represents a complex filamentary wire implant.
CONCLUSIONS CONCLUSIONS
The proposed simplified methodology, suitable to be applied to closed loop wires in the low to intermediate frequency range, is found to be sufficiently accurate and easy to apply in realistic exposure scenarios. This modeling tool allows analyzing different types of small implants, from coronary and biliary duct stents to orthopedic grids, under a variety of exposure scenarios.

Identifiants

pubmed: 36566651
pii: S0169-2607(22)00697-6
doi: 10.1016/j.cmpb.2022.107316
pii:
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

107316

Informations de copyright

Copyright © 2022. Published by Elsevier B.V.

Déclaration de conflit d'intérêts

Declaration of Competing Interest All Authors declare that they have no conflict of interest.

Auteurs

Oriano Bottauscio (O)

Istituto Nazionale di Ricerca Metrologica (INRiM), Strada Delle Cacce 91, 10135 Torino, Italy. Electronic address: o.bottauscio@inrim.it.

Alessandro Arduino (A)

Istituto Nazionale di Ricerca Metrologica (INRiM), Strada Delle Cacce 91, 10135 Torino, Italy.

Mario Chiampi (M)

Istituto Nazionale di Ricerca Metrologica (INRiM), Strada Delle Cacce 91, 10135 Torino, Italy.

Luca Zilberti (L)

Istituto Nazionale di Ricerca Metrologica (INRiM), Strada Delle Cacce 91, 10135 Torino, Italy.

Articles similaires

1.00
Humans Skull Infant Child, Preschool Infant, Newborn
Humans Heart Atria Pulmonary Veins Stenosis, Pulmonary Vein Stents

Personalized bioceramic grafts for craniomaxillofacial bone regeneration.

Ana Beatriz G de Carvalho, Maedeh Rahimnejad, Rodrigo L M S Oliveira et al.
1.00
Humans Bone Regeneration Ceramics Printing, Three-Dimensional Tissue Scaffolds
Animals Biological Evolution Skull Perciformes X-Ray Microtomography

Classifications MeSH