It's not only the pump: Assessment of human factors of wearable components and user experience of patients with left ventricular assist devices.
design
human factors
human factors engineering
left ventricular assist device
mechanical circulatory support
usability
wearables
Journal
The Journal of heart and lung transplantation : the official publication of the International Society for Heart Transplantation
ISSN: 1557-3117
Titre abrégé: J Heart Lung Transplant
Pays: United States
ID NLM: 9102703
Informations de publication
Date de publication:
04 2023
04 2023
Historique:
received:
19
08
2022
revised:
24
10
2022
accepted:
18
12
2022
medline:
29
3
2023
pubmed:
23
1
2023
entrez:
22
1
2023
Statut:
ppublish
Résumé
Despite design improvements in left ventricular assist devices (LVADs) over the past decade, limitations of external, wearable VAD components affect patient quality of life and safety. The aim of this study was to describe both user experience and human factor issues of 2 contemporary LVADs. This single-center, cross-sectional study included LVAD outpatients who were at least 3 months after implantation. Before developing the 16-item survey, a systematic literature review and 2-round Delphi method involving 9 VAD clinicians were used to select items in 6 domains: power supply, emergency situations, wearability, mobility, and freedom to travel, user modifications, lifestyle, and home adaptations. Fifty-eight patients (61.6 ± 11.6 years, 13.8% female, HeartMate 3 (HM3)/HVAD: n = 39/19) completed the one-time survey after median of 853 days on device: 10.3% reported problems changing power supply, 12.7% unintentional driveline disconnection (HM3: 5.6% vs HVAD: 26.3%, p = 0.041). Against the recommendation 74.1% sleep with battery-support (HM3: 88.9% vs HVAD: 44.4%, p = 0.001). About 65.3% criticized the carry bag weight/size (HM3: 71.4% vs HVAD: 50.0%, p = 0.035), thus 24.1% wear an own carrying-system, 42.1% modified their wearables, 38.9% their clothing, and 65.3% their home to cope with life on LVAD support. Mobility is reduced due to limited wearability: 18.9% went abroad (only 3.7% by plane) and 40.0% use less public transport than before implantation (the older the less: r = -0.37, p = 0.013). HVAD and HM3 wearables still show a variety of human factors issues and potential for improved user experience. User-centered design and incorporation of patient feedback may increase user satisfaction, and patient safety.
Sections du résumé
BACKGROUND
Despite design improvements in left ventricular assist devices (LVADs) over the past decade, limitations of external, wearable VAD components affect patient quality of life and safety. The aim of this study was to describe both user experience and human factor issues of 2 contemporary LVADs.
METHODS
This single-center, cross-sectional study included LVAD outpatients who were at least 3 months after implantation. Before developing the 16-item survey, a systematic literature review and 2-round Delphi method involving 9 VAD clinicians were used to select items in 6 domains: power supply, emergency situations, wearability, mobility, and freedom to travel, user modifications, lifestyle, and home adaptations.
RESULTS
Fifty-eight patients (61.6 ± 11.6 years, 13.8% female, HeartMate 3 (HM3)/HVAD: n = 39/19) completed the one-time survey after median of 853 days on device: 10.3% reported problems changing power supply, 12.7% unintentional driveline disconnection (HM3: 5.6% vs HVAD: 26.3%, p = 0.041). Against the recommendation 74.1% sleep with battery-support (HM3: 88.9% vs HVAD: 44.4%, p = 0.001). About 65.3% criticized the carry bag weight/size (HM3: 71.4% vs HVAD: 50.0%, p = 0.035), thus 24.1% wear an own carrying-system, 42.1% modified their wearables, 38.9% their clothing, and 65.3% their home to cope with life on LVAD support. Mobility is reduced due to limited wearability: 18.9% went abroad (only 3.7% by plane) and 40.0% use less public transport than before implantation (the older the less: r = -0.37, p = 0.013).
CONCLUSIONS
HVAD and HM3 wearables still show a variety of human factors issues and potential for improved user experience. User-centered design and incorporation of patient feedback may increase user satisfaction, and patient safety.
Identifiants
pubmed: 36682893
pii: S1053-2498(22)02269-0
doi: 10.1016/j.healun.2022.12.015
pii:
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
466-477Informations de copyright
Copyright © 2022 The Authors. Published by Elsevier Inc. All rights reserved.