Repeatability and Reproducibility Uncertainty in Magnetic Resonance-Based Electric Properties Tomography of a Homogeneous Phantom.
electric properties tomography
magnetic resonance imaging
phantoms
quantitative imaging
uncertainty
Journal
Tomography (Ann Arbor, Mich.)
ISSN: 2379-139X
Titre abrégé: Tomography
Pays: Switzerland
ID NLM: 101671170
Informations de publication
Date de publication:
17 02 2023
17 02 2023
Historique:
received:
28
01
2023
revised:
10
02
2023
accepted:
14
02
2023
entrez:
24
2
2023
pubmed:
25
2
2023
medline:
3
3
2023
Statut:
epublish
Résumé
Uncertainty assessment is a fundamental step in quantitative magnetic resonance imaging because it makes comparable, in a strict metrological sense, the results of different scans, for example during a longitudinal study. Magnetic resonance-based electric properties tomography (EPT) is a quantitative imaging technique that retrieves, non-invasively, a map of the electric properties inside a human body. Although EPT has been used in some early clinical studies, a rigorous experimental assessment of the associated uncertainty has not yet been performed. This paper aims at evaluating the repeatability and reproducibility uncertainties in phase-based Helmholtz-EPT applied on homogeneous phantom data acquired with a clinical 3 T scanner. The law of propagation of uncertainty is used to evaluate the uncertainty in the estimated conductivity values starting from the uncertainty in the acquired scans, which is quantified through a robust James-Stein shrinkage estimator to deal with the dimensionality of the problem. Repeatable errors are detected in the estimated conductivity maps and are quantified for various values of the tunable parameters of the EPT implementation. The spatial dispersion of the estimated electric conductivity maps is found to be a good approximation of the reproducibility uncertainty, evaluated by changing the position of the phantom after each scan. The results underpin the use of the average conductivity (calculated by weighting the local conductivity values by their uncertainty and taking into account the spatial correlation) as an estimate of the conductivity of the homogeneous phantom.
Identifiants
pubmed: 36828386
pii: tomography9010034
doi: 10.3390/tomography9010034
pmc: PMC9961522
doi:
Types de publication
Journal Article
Research Support, Non-U.S. Gov't
Langues
eng
Sous-ensembles de citation
IM
Pagination
420-435Références
IEEE Trans Biomed Eng. 2021 Jan;68(1):236-246
pubmed: 32365014
Phys Med Biol. 1996 Nov;41(11):2251-69
pubmed: 8938025
IEEE Trans Med Imaging. 2015 Sep;34(9):1788-96
pubmed: 25706578
Magn Reson Med. 2012 Feb;67(2):552-61
pubmed: 21710613
Magn Reson Med. 2007 Jan;57(1):192-200
pubmed: 17191242
Insights Imaging. 2013 Apr;4(2):147-52
pubmed: 23397519
Eur Radiol. 2018 Jan;28(1):348-355
pubmed: 28698943
Stat Appl Genet Mol Biol. 2007;6:Article9
pubmed: 17402924
Comput Math Methods Med. 2013;2013:546562
pubmed: 23573170
Diagnostics (Basel). 2021 Jan 26;11(2):
pubmed: 33530587
Magn Reson Med. 1996 Feb;35(2):246-51
pubmed: 8622590
Magn Reson Med. 2022 Mar;87(3):1500-1506
pubmed: 34739149
J Magn Reson Imaging. 2015 Aug;42(2):371-8
pubmed: 25413153
IEEE Trans Med Imaging. 2015 Nov;34(11):2220-32
pubmed: 25955582
Magn Reson Med. 2012 Oct;68(4):1117-26
pubmed: 22213053
Magn Reson Med. 2012 Nov;68(5):1517-26
pubmed: 22252850
IEEE Trans Med Imaging. 2015 Feb;34(2):541-50
pubmed: 25312919
J Magn Reson Imaging. 2021 Nov;54(5):1689-1691
pubmed: 33998080
IEEE Trans Med Imaging. 2012 Feb;31(2):430-7
pubmed: 21990329
Magn Reson Med. 2018 Jul;80(1):90-100
pubmed: 29144031
IEEE Trans Med Imaging. 2018 Sep;37(9):2080-2089
pubmed: 29994520
Magn Reson Med. 2011 Aug;66(2):456-66
pubmed: 21773985
Magn Reson Med. 2010 May;63(5):1315-22
pubmed: 20432302
Int J Hyperthermia. 2020;37(1):992-1007
pubmed: 32806979
Stat Appl Genet Mol Biol. 2005;4:Article32
pubmed: 16646851
IEEE Trans Med Imaging. 2017 Jul;36(7):1396-1404
pubmed: 28186884
Br J Radiol. 2021 Apr 1;94(1120):20201215
pubmed: 33710907
Adv Exp Med Biol. 2022;1395:23-27
pubmed: 36527608
IEEE Trans Biomed Eng. 2018 Feb;65(2):282-293
pubmed: 29035208
IEEE Trans Biomed Eng. 2017 Nov;64(11):2515-2530
pubmed: 28829299
Eur Radiol. 2016 Jul;26(7):2317-26
pubmed: 26497503
IEEE Trans Biomed Eng. 2020 Jan;67(1):3-15
pubmed: 30908189