Optimal Degree of Balloon-Expandable Transcatheter Valve Oversizing in Patients With Borderline Aortic Annulus Measurements: Insights From a Multicenter Real-World Experience.
hemodynamics
outcomes
oversizing
paravalvular aortic regurgitation
permanent pacemaker implantation
Journal
Circulation. Cardiovascular interventions
ISSN: 1941-7632
Titre abrégé: Circ Cardiovasc Interv
Pays: United States
ID NLM: 101499602
Informations de publication
Date de publication:
03 2023
03 2023
Historique:
entrez:
21
3
2023
pubmed:
22
3
2023
medline:
24
3
2023
Statut:
ppublish
Résumé
The potential benefit of using larger or smaller transcatheter heart valves (THV) in patients with borderline aortic annulus measurement (BAM) remains uncertain. The objective of this study was to evaluate the clinical outcomes associated with the selection of larger or smaller THV in the context of BAM. This was a multicenter observational study including patients who underwent transcatheter aortic valve replacement with the SAPIEN 3 or SAPIEN 3 Ultra-valve systems (Edwards Lifesciences, Irvine, CA) from April 2014 to June 2021. BAM was defined according to the manufacturer sizing chart and included the following annulus areas: 314 to 346, 400 to 430, 500 to 546 mm From a total of 2467 patients, BAM was identified in 852 patients (34.5%). A larger and smaller THV was selected in 338 (39.7%) and 514 patients (60.3%) patients, respectively. The choice of a larger THV was associated (before and after propensity matching) with a higher risk of new-onset left bundle branch block (HR, 2.25 [95% CI, 1.39-3.65; In patient with BAM, selecting a larger SAPIEN 3/Ultra THV increased the risk of conduction disturbances without any benefit on valve hemodynamics and clinical outcomes.
Sections du résumé
BACKGROUND
The potential benefit of using larger or smaller transcatheter heart valves (THV) in patients with borderline aortic annulus measurement (BAM) remains uncertain. The objective of this study was to evaluate the clinical outcomes associated with the selection of larger or smaller THV in the context of BAM.
METHODS
This was a multicenter observational study including patients who underwent transcatheter aortic valve replacement with the SAPIEN 3 or SAPIEN 3 Ultra-valve systems (Edwards Lifesciences, Irvine, CA) from April 2014 to June 2021. BAM was defined according to the manufacturer sizing chart and included the following annulus areas: 314 to 346, 400 to 430, 500 to 546 mm
RESULTS
From a total of 2467 patients, BAM was identified in 852 patients (34.5%). A larger and smaller THV was selected in 338 (39.7%) and 514 patients (60.3%) patients, respectively. The choice of a larger THV was associated (before and after propensity matching) with a higher risk of new-onset left bundle branch block (HR, 2.25 [95% CI, 1.39-3.65;
CONCLUSIONS
In patient with BAM, selecting a larger SAPIEN 3/Ultra THV increased the risk of conduction disturbances without any benefit on valve hemodynamics and clinical outcomes.