Evaluation of surface image guidance and Deep inspiration Breath Hold technique for breast treatments with Halcyon.
Catalyst Surface Guided Therapy (SGRT)
Deep Inspiration Breath Hold (DIBH)
Halcyon
Journal
Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)
ISSN: 1724-191X
Titre abrégé: Phys Med
Pays: Italy
ID NLM: 9302888
Informations de publication
Date de publication:
Apr 2023
Apr 2023
Historique:
received:
23
11
2021
revised:
13
12
2022
accepted:
14
03
2023
medline:
7
4
2023
pubmed:
30
3
2023
entrez:
29
3
2023
Statut:
ppublish
Résumé
To evaluate the accuracy/agreement of a three-camera Catalyst Surface Guided Radiation Therapy (SGRT) system on a closed-gantry Halcyon for Free-Breathing (FB) and Deep Inspiration Breath Hold (DIBH) breast-only treatments. The SGRT positioning agreement with Halcyon couch and cone-beam computed tomography (CBCT) was evaluated on phantom and by evaluation of 2401 FB and 855 DIBH breast-only treatment sessions. The DIBH agreement was evaluated using a programmable moving support. Dose agreement was evaluated for manual SGRT-assisted beam interruption and Halcyon arc beam interruption. Geometrical phantom agreement was < 0.4 mm. Couch and SGRT agreement for an anthropomorphic phantom resulted in 95% limits of agreement in Right-Left/Feet-Head/Posterior-Anterior (RL/FH/PA) directions of respectively ± 0.4/0.8/0.5 mm and ± 1.1/1.1/0.6 mm in the virtual and real isocenter. FB-SGRT-assisted patient positioning compared to CBCT positioning resulted in RL/FH/PA systematic differences of -0.1/0.1/2.0 mm with standard deviations of 2.7/2.8/2.4 mm. This mean systematic difference had three origins: a) couch sag/isocenter difference of ≤ 0.5 mm. b) Average reconstructed FB-CBCT images do not visually represent the average respiratory position. c) CBCT-based positioning focused on the inner thoracic interface, which can introduce a mean positioning difference between SGRT and CBCT. Manual SGRT-assisted beam interruption and arc interruptions resulted in mean gamma passing rates > 97% (0.5%/0.5 mm) and mean absolute differences < 0.3%. Accuracy was comparable with breast-only C-arm SGRT techniques, with different tradeoffs. Depending on the patient's morphology, real-time tracking accuracy in the real isocenter can be reduced. This study demonstrates possible discordances between SGRT and CBCT positioning for breast.
Identifiants
pubmed: 36989980
pii: S1120-1797(23)00041-8
doi: 10.1016/j.ejmp.2023.102564
pii:
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
102564Informations de copyright
Copyright © 2023 Associazione Italiana di Fisica Medica e Sanitaria. Published by Elsevier Ltd. All rights reserved.
Déclaration de conflit d'intérêts
Declaration of Competing Interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.