Clinical utility of single-shot echo-planar diffusion-weighted imaging using L1-regularized iterative sensitivity encoding in prostate MRI.


Journal

Medicine
ISSN: 1536-5964
Titre abrégé: Medicine (Baltimore)
Pays: United States
ID NLM: 2985248R

Informations de publication

Date de publication:
25 Apr 2023
Historique:
medline: 1 5 2023
pubmed: 28 4 2023
entrez: 28 4 2023
Statut: ppublish

Résumé

We investigated the ability of echo-planar imaging with L1-regularized iterative sensitivity encoding-based diffusion-weighted imaging (DWI) to improve the image quality and reduce the scanning time in prostate magnetic resonance imaging. We retrospectively analyzed 109 cases of prostate magnetic resonance imaging. We compared variables in the quantitative and qualitative assessments among 3 imaging groups: conventional parallel imaging-based DWI (PI-DWI) with an acquisition time of 3 minutes 15 seconds; echo-planar imaging with L1-regularized iterative sensitivity encoding-based DWI (L1-DWI) with a normal acquisition time (L1-DWINEX12) of 3 minutes 15 seconds; and L1-DWI with a half acquisition time (L1-DWINEX6) of 1 minute 45 seconds. As a quantitative assessment, the signal-to-noise ratio (SNR) of DWI (SNR-DWI), the contrast-to-noise ratio (CNR) of DWI (CNR-DWI), and the CNR of apparent diffusion coefficient were measured. As a qualitative assessment, the image quality and visual detectability of prostate carcinoma were evaluated. In the quantitative analysis, L1-DWINEX12 showed significantly higher SNR-DWI than PI-DWI (P = .0058) and L1-DWINEX6 (P < .0001). In the qualitative analysis, the image quality score for L1-DWINEX12 was significantly higher than those of PI-DWI and L1-DWINEX6. A non-inferiority assessment demonstrated that L1-DWINEX6 was non-inferior to PI-DWI in terms of both quantitative CNR-DWI and qualitative grading of image quality with a <20% inferior margin. L1-DWI successfully demonstrated a reduced scanning time while maintaining good image quality.

Identifiants

pubmed: 37115048
doi: 10.1097/MD.0000000000033639
pii: 00005792-202304250-00007
pmc: PMC10146059
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

e33639

Informations de copyright

Copyright © 2023 the Author(s). Published by Wolters Kluwer Health, Inc.

Déclaration de conflit d'intérêts

The authors declare relationships with the following companies: MY is currently employed by Philips Japan. The remaining authors have no conflicts of interest to disclose.

Références

Hoeks CM, Barentsz JO, Hambrock T, et al. Prostate cancer: multiparametric MR imaging for detection, localization, and staging. Radiology. 2011;261:46–66.
Weinreb JC, Barentsz JO, Choyke PL, et al. PI-RADS prostate imaging – reporting and data system: 2015, version 2. Eur Urol. 2016;69:16–40.
Turkbey B, Rosenkrantz AB, Haider MA, et al. Prostate imaging reporting and data system version 2.1: 2019 update of prostate imaging reporting and data system version 2. Eur Urol. 2019;76:340–51.
Akisik FM, Sandrasegaran K, Aisen AM, et al. Abdominal MR imaging at 3.0 T. Radiographics. 2007;27:1433–44.
Noël P, Bammer R, Reinhold C, et al. Parallel imaging artifacts in body magnetic resonance imaging. Can Assoc Radiol J. 2009;60:91–8.
Golay X, de Zwart JA, Ho YC, et al. Parallel imaging techniques in functional MRI. Top Magn Reson Imaging. 2004;15:255–65.
Liang D, Liu B, Wang J, et al. Accelerating SENSE using compressed sensing. Magn Reson Med. 2009;62:1574–84.
Lustig M, Donoho D, Pauly JM. Sparse MRI: the application of compressed sensing for rapid MR imaging. Magn Reson Med. 2007;58:1182–95.
Song JS, Kim SH, Kuehn B, et al. Optimized breath-hold compressed-sensing 3D MR cholangiopancreatography at 3T: image quality analysis and clinical feasibility assessment. Diagnostics (Basel). 2020;10:376.
Jiang Y, Wang X, Zhu L, et al. Compressed-sensing accelerated magnetic resonance imaging of inner ear. J Appl Clin Med Phys. 2021;22:332–8.
Meister RL, Groth M, Jürgens JHW, et al. Compressed SENSE in pediatric brain tumor MR imaging: assessment of image quality, examination time and energy release. Clin Neuroradiol. 2022;32:725–33.
Yoneyama M, Morita K, Peeters J, et al. Noise Reduction in Prostate Single-Shot DW-EPI utilizing Compressed SENSE Framework. Montreal, QC, Canada: ISMRM. 2019.
Morita K, Yoneyama M, Nakaura T, et al. Pseudo-3D Diffusion-Weighted Imaging of the Brain using Echo Planar Imaging with Compressed SENSE (EPICS). Montreal, QC. Canada: ISMRM. 2019.
Kaga T, Noda Y, Mori T, et al. Diffusion-weighted imaging of the abdomen using echo planar imaging with compressed SENSE: feasibility, image quality, and ADC value evaluation. Eur J Radiol. 2021;142:109889.
Tamada T, Ueda Y, Kido A, et al. Clinical application of single-shot echo-planar diffusion-weighted imaging with compressed SENSE in prostate MRI at 3T: preliminary experience. MAGMA. 2022;35:549–56.
Yoshida N, Nakaura T, Morita K, et al. Echo planar imaging with compressed sensitivity encoding (EPICS): usefulness for head and neck diffusion-weighted MRI. Eur J Radiol. 2022;155:110489.
Klingebiel M, Ullrich T, Quentin M, et al. Advanced diffusion weighted imaging of the prostate: comparison of readout-segmented multi-shot, parallel-transmit and single-shot echo-planar imaging. Eur J Radiol. 2020;130:109161.
Epstein JI, Egevad L, Amin MB, et al. The 2014 International Society of Urological Pathology (ISUP) consensus conference on gleason grading of prostatic carcinoma: definition of grading patterns and proposal for a new grading system. Am J Surg Pathol. 2016;40:244–52.
Park IK, Park J, Kim TH, et al. Non-inferior low-dose coronary computed tomography angiography image quality with knowledge-based iterative model reconstruction for overweight patients. PLoS One. 2018;13:e0209243.
Lawrence EM, Zhang Y, Starekova J, et al. Reduced field-of-view and multi-shot DWI acquisition techniques: prospective evaluation of image quality and distortion reduction in prostate cancer imaging. Magn Reson Imaging. 2022;93:108–14.
Ueda T, Ohno Y, Yamamoto K, et al. Deep learning reconstruction of diffusion-weighted MRI improves image quality for prostatic imaging. Radiology. 2022;303:373–81.

Auteurs

Noriko Nishioka (N)

Department of Diagnostic and Interventional Radiology, Hokkaido University Hospital, Kita-Ku, Sapporo, Japan.
Department of Diagnostic Imaging, Faculty of Medicine and Graduate School of Medicine, Hokkaido University, Kita-Ku, Sapporo, Japan.

Noriyuki Fujima (N)

Department of Diagnostic and Interventional Radiology, Hokkaido University Hospital, Kita-Ku, Sapporo, Japan.

Satonori Tsuneta (S)

Department of Diagnostic and Interventional Radiology, Hokkaido University Hospital, Kita-Ku, Sapporo, Japan.

Masami Yoneyama (M)

Philips Japan, Minato-ku, Tokyo, Japan.

Ryuji Matsumoto (R)

Department of Renal and Genitourinary Surgery, Faculty of Medicine and Graduate School of Medicine, Hokkaido University, Kita-Ku, Sapporo, Japan.

Takashige Abe (T)

Department of Renal and Genitourinary Surgery, Faculty of Medicine and Graduate School of Medicine, Hokkaido University, Kita-Ku, Sapporo, Japan.

Rina Kimura (R)

Department of Diagnostic and Interventional Radiology, Hokkaido University Hospital, Kita-Ku, Sapporo, Japan.
Department of Diagnostic Imaging, Faculty of Medicine and Graduate School of Medicine, Hokkaido University, Kita-Ku, Sapporo, Japan.

Keita Sakamoto (K)

Department of Diagnostic and Interventional Radiology, Hokkaido University Hospital, Kita-Ku, Sapporo, Japan.

Fumi Kato (F)

Department of Diagnostic and Interventional Radiology, Hokkaido University Hospital, Kita-Ku, Sapporo, Japan.

Kohsuke Kudo (K)

Department of Diagnostic and Interventional Radiology, Hokkaido University Hospital, Kita-Ku, Sapporo, Japan.
Department of Diagnostic Imaging, Faculty of Medicine and Graduate School of Medicine, Hokkaido University, Kita-Ku, Sapporo, Japan.
Department of Advanced Diagnostic Imaging Development, Faculty of Medicine and Graduate School of Medicine, Hokkaido University, Kita-Ku, Sapporo, Japan.
Global Center for Biomedical Science and Engineering, Faculty of Medicine, Hokkaido University, Kita-Ku, Sapporo, Hokkaido, Japan.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH