Refurbishment of Extracorporeal Life Support Oxygenators in the Context of In Vitro Testing.


Journal

ASAIO journal (American Society for Artificial Internal Organs : 1992)
ISSN: 1538-943X
Titre abrégé: ASAIO J
Pays: United States
ID NLM: 9204109

Informations de publication

Date de publication:
01 10 2023
Historique:
medline: 2 10 2023
pubmed: 14 6 2023
entrez: 14 6 2023
Statut: ppublish

Résumé

Refurbishing single use extracorporeal membrane oxygenation (ECMO) oxygenators for in vitro research applications is common. However, the refurbishment protocols that are established in respective laboratories have never been evaluated. In the present study, we aim at proving the relevance of a well-designed refurbishing protocol by quantifying the burden of repeatedly reused oxygenators. We used the same three oxygenators in 5 days of 6 hours whole blood experiments. During each experiment day, the performance of the oxygenators was measured through the evaluation of gas transfer. Between experiment days, each oxygenator was refurbished applying three alternative refurbishment protocols based on purified water, pepsin and citric acid, and hydrogen peroxide solutions, respectively. After the last experiment day, we disassembled the oxygenators for visual inspection of the fiber mats. The refurbishment protocol based on purified water showed strong degeneration with a 40-50 %-performance drop and clearly visible debris on the fiber mats. Hydrogen peroxide performed better; nevertheless, it suffered a 20% decrease in gas transfer as well as clearly visible debris. Pepsin/citric acid performed best in the field, but also suffered from 10% performance loss and very few, but visible debris. The study showed the relevance of a well-suited and well-designed refurbishment protocol. The distinct debris on the fiber mats also suggests that reusing oxygenators is ill-advised for many experiment series, especially regarding hemocompatibility and in vivo testing. Most of all, this study revealed the relevance of stating the status of test oxygenators and, if refurbished, comment on the implemented refurbishment protocol in detail.

Identifiants

pubmed: 37314830
doi: 10.1097/MAT.0000000000001999
pii: 00002480-990000000-00259
doi:

Substances chimiques

Hydrogen Peroxide BBX060AN9V
Pepsin A EC 3.4.23.1
Citric Acid 2968PHW8QP
Water 059QF0KO0R

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

924-931

Informations de copyright

Copyright © ASAIO 2023.

Déclaration de conflit d'intérêts

Disclosure: The authors have no conflicts of interest to report.

Références

Arens J, Schnoering H, Pfennig M, et al.: The Aachen MiniHLM--a miniaturized heart-lung machine for neonates with an integrated rotary blood pump. Artif Organs 34: 707–713, 2010.
Baud FJ, Wasram Jetha-Jamal T, Vicca S, Raphalen J-H, Lamhaut L: Disposition of gentamicin and amikacin in extracorporeal membrane oxygenation using a heparin-coated filter: An in vitro assessment. Int J Artif Organs 45: 588–592, 2022.
Karagiannidis C, Hesselmann F, Fan E: Physiological and technical considerations of extracorporeal CO 2 removal. Crit Care 23: 75, 2019.
Evseev AK, Zhuravel SV, Alentiev AY, et al.: Membranes in extracorporeal blood oxygenation technology. Membr Technol 1: 201–211, 2019.
Combes A, Peek GJ, Hajage D, et al.: ECMO for severe ARDS: Systematic review and individual patient data meta-analysis. Intensive Care Med 46: 2048–2057, 2020.
Makdisi G, Wang I-w: Extra corporeal membrane oxygenation (ECMO) review of a lifesaving technology. J Thorac Dis 7: E166–E176, 2015.
Schmidt M, Burrell A, Roberts L, et al.: Predicting survival after ECMO for refractory cardiogenic shock: The survival after veno-arterial-ECMO (SAVE)-score. Eur Heart J 36: 2246–2256, 2015.
Stub D, Bernard S, Pellegrino V, et al.: Refractory cardiac arrest treated with mechanical CPR, hypothermia, ECMO and early reperfusion (the CHEER trial). Resuscitation 86: 88–94, 2015.
Ramanathan K, Antognini D, Combes A, et al.: Planning and provision of ECMO services for severe ARDS during the COVID-19 pandemic and other outbreaks of emerging infectious diseases. Lancet Respirat Med 8: 518–526, 2020.
Chan CHH, Ki KK, Zhang M, et al.: Extracorporeal membrane oxygenation-induced hemolysis: An in vitro study to appraise causative factors. Membranes (Basel) 11: 313, 2021.
Orizondo RA, May AG, Madhani SP, et al.: In vitro characterization of the Pittsburgh pediatric ambulatory lung. ASAIO J 64: 806–811, 2018.
Nasr VG, Meserve J, Pereira LM, et al.: Sedative and analgesic drug sequestration after a single bolus injection in an ex vivo extracorporeal membrane oxygenation infant circuit. ASAIO J 65: 187–191, 2019.
Wang S, Moroi MK, Kunselman AR, Myers JL, Ündar A: Evaluation of centrifugal blood pumps in term of hemodynamic performance using simulated neonatal and pediatric ECMO circuits. Artif Organs 44: 16–27, 2020.
Zhang T, Cheng G, Koert A, et al.: Functional and biocompatibility performances of an integrated Maglev pump-oxygenator. Artif Organs 33: 36–45, 2009.
International Organization for Standardization: Cardiovascular Implants and Artificial Organs: Cardiovascular Implants and Artificial Organs — Blood-Gas Exchangers (Oxygenators);11.040.40 Implants for Surgery, Prosthetics and Orthotics(7199:2016). 2016. https://www.beuth.de/en/standard/din-en-iso-7199/317329787 . Accessed March 21, 2021.
Schraven L, Kaesler A, Flege C, et al.: Effects of pulsatile blood flow on oxygenator performance. Artif Organs 42: 410–419, 2018.
Arens J, Schraven L, Kaesler A, et al.: Development and evaluation of a variable, miniaturized oxygenator for various test methods. Artif Organs 47: 695−704, 2023.
Arens J, Grottke O, Haverich A, et al.: Toward a long-term artificial lung. ASAIO J 66: 847–854, 2020.
Hesselmann F, Scherenberg N, Bongartz P, et al.: Structure-dependent gas transfer performance of 3D-membranes for artificial membrane lungs. J Membr Sci 2021: 71, 1193.
Wiegmann B, Seggern H von, Höffler K, et al.: Developing a biohybrid lung - sufficient endothelialization of poly-4-methly-1-pentene gas exchange hollow-fiber membranes. J Mech Behav Biomed Mater 60: 301–311, 2016.
Thompson AJ, Buchan S, Carr B, et al.: Low-resistance, concentric-gated pediatric artificial lung for end-stage lung failure. ASAIO J 66: 423–432, 2020.
Swol J, Shigemura N, Ichiba S, Steinseifer U, Anraku M, Lorusso R: Artificial lungs--where are we going with the lung replacement therapy? Artif Organs 44: 1135–1149, 2020.
Strudthoff LJ, Lüken H, Jansen SV, et al.: In vitro and in vivo feasibility study for a portable VV-ECMO and ECCO2R system. Membranes (Basel) 12: 133, 2022.
Orizondo RA, Cardounel AJ, Kormos R, et al.: Artificial lungs: Current status and future directions. Curr Transpl Rep 6: 307–315, 2019.
Ontaneda A, Annich GM: Novel surfaces in extracorporeal membrane oxygenation circuits. Front Med (Lausanne) 5: 321, 2018.
Nichols JE, La Francesca S, Niles JA, et al.: Production and transplantation of bioengineered lung into a large-animal model. Sci Transl Med 10: eaao3926, 2018.
Klein S, Hesselmann F, Djeljadini S, et al.: EndOxy: Dynamic long-term evaluation of endothelialized gas exchange membranes for a biohybrid lung. Ann Biomed Eng 48: 747–756, 2020.
Custer JR: The evolution of patient selection criteria and indications for extracorporeal life support in pediatric cardiopulmonary failure: Next time, let’s not eat the bones. Organogenesis 7: 13–22, 2011.

Auteurs

Lasse J Strudthoff (LJ)

From the Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Medical Faculty, RWTH Aachen University, Aachen, Germany.

Felix Hesselmann (F)

From the Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Medical Faculty, RWTH Aachen University, Aachen, Germany.
Enmodes GmbH, Aachen, Germany.

Johanna C Clauser (JC)

From the Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Medical Faculty, RWTH Aachen University, Aachen, Germany.

Jutta Arens (J)

From the Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Medical Faculty, RWTH Aachen University, Aachen, Germany.

Articles similaires

Animals Dietary Fiber Dextran Sulfate Mice Disease Models, Animal
Silicon Dioxide Water Hot Temperature Compressive Strength X-Ray Diffraction
1.00
Oryza Agricultural Irrigation Potassium Sodium Soil
Humans Out-of-Hospital Cardiac Arrest Female Male Japan

Classifications MeSH