Implementation and evaluation of the AAPM TG 111 for radiation dosimetry in a state-of-the-art C-arm cone beam CT system.


Journal

Journal of radiological protection : official journal of the Society for Radiological Protection
ISSN: 1361-6498
Titre abrégé: J Radiol Prot
Pays: England
ID NLM: 8809257

Informations de publication

Date de publication:
28 09 2023
Historique:
received: 14 07 2023
accepted: 18 09 2023
medline: 23 10 2023
pubmed: 19 9 2023
entrez: 18 9 2023
Statut: epublish

Résumé

The aim of this study was to assess the dosimetric characteristics of a state-of-the-art C-arm cone beam computed tomography (CBCT) system using the methodology proposed by the American Association of Physicists in Medicine (AAPM) Task Group (TG) 111. The dose measurement methodology described in AAPM TG 111 for wide cone beam acquisitions without table translation was employed to estimate equilibrium beam length (αeq‾) and equilibrium dose (feq‾) in various interventional task-specific protocols with different tube arc projection geometries. Dose profiles were derived from point dose measurements in the centre and peripheral locations of the ICRU/AAPM and standard polymethyl methacrylate (PMMA) body phantom. Dose measurements were performed in phantom and free-in-air using a solid-state point detector. Monte Carlo (MC) based simulation dosimetry was performed to quantify the inhomogeneous dose patterns imparted in the phantoms. Estimatedαeq‾andfeq‾on the ICRU/AAPM phantom was up to 49.4 cm and 6.17 mGy/100 mAs, respectively. Corresponding values determined on the PMMA phantom were 139 cm and 8.8 mGy/100 mAs, respectively. Free-in-air dose measurement ranged from 1.43 mGy/100 mAs to 5.93 mGy/100 mAs. Per cent difference inαeq‾andfeq‾between MC simulation and solid-state point detector measurement methods in the ICRU/AAPM phantom were within 16% and 18%, respectively. Manufacturers can use the presented methodology to characterize the dosimetric properties of C-arm CBCT systems. Clinical medical physicists may follow this methodology to verify corresponding data provided by the manufacturer and check for C-arm CBCT system performance dosimetric consistency.

Identifiants

pubmed: 37722379
doi: 10.1088/1361-6498/acfaa9
doi:

Substances chimiques

Polymethyl Methacrylate 9011-14-7

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Informations de copyright

© 2023 Society for Radiological Protection. Published on behalf of SRP by IOP Publishing Limited. All rights reserved.

Auteurs

Antonios E Papadakis (AE)

University Hospital of Heraklion, Department of Medical Physics, Heraklion, Crete, Greece.

John Damilakis (J)

University of Crete, School of Medicine, Department of Medical Physics, Heraklion, Crete, Greece.

Articles similaires

Humans Magnetic Resonance Imaging Phantoms, Imaging Infant, Newborn Signal-To-Noise Ratio
Humans Meta-Analysis as Topic Sample Size Models, Statistical Computer Simulation
Humans Algorithms Software Artificial Intelligence Computer Simulation
Humans Robotic Surgical Procedures Clinical Competence Male Female

Classifications MeSH