Monte Carlo investigation of dose distribution of uniformly and non-uniformly loaded standard and notched eye plaques.
Monte Carlo dosimetry
eye plaque brachytherapy
non-uniform loading
notched eye plaques
Journal
Journal of applied clinical medical physics
ISSN: 1526-9914
Titre abrégé: J Appl Clin Med Phys
Pays: United States
ID NLM: 101089176
Informations de publication
Date de publication:
Dec 2023
Dec 2023
Historique:
revised:
06
08
2023
received:
08
06
2023
accepted:
16
08
2023
medline:
4
12
2023
pubmed:
22
9
2023
entrez:
22
9
2023
Statut:
ppublish
Résumé
To investigate the effect of using non-uniform loading and notched plaques on dose distribution for eye plaques. Using EGSnrc Monte Carlo (MC) simulations, we investigate eye plaque dose distributions in water and in an anatomically representative eye phantom. Simulations were performed in accordance with TG43 formalism and compared against full MC simulations which account for inter-seed and inhomogeneity effects. For standard plaque configurations, uniformly and non-uniformly loaded plaque dose distributions in water showed virtually no difference between each other. For standard plaque, the MC calculated dose distribution in planes parallel to the plaque is narrower than the TG43 calculation due to attenuation at the periphery of the plaque by the modulay. MC calculated the dose behind the plaque is fully attenuated. Similar results were found for the notched plaque, with asymmetric attenuation along the plane of the notch. Cumulative dose volume histograms showed significant reductions in the calculated MC doses for both tumor and eye structures, compared to TG43 calculations. The effect was most pronounced for the notch plaque where the MC dose to the optic nerve was greatly attenuated by the modulay surrounding the optic nerve compared to the TG43. Thus, a reduction of optic nerve D95% from 14 to 0.2 Gy was observed, when comparing the TG43 calculation to the MC result. The tumor D95% reduced from 89.2 to 79.95 Gy for TG43 and MC calculations, respectively. TG43 calculations overestimate the absolute dose and the lateral dose distribution of both standard and notched eye plaques, leading to the dose overestimation for the target and organs at risk. The dose matching along the central axis for the non-uniformly loaded plaques to that of uniformly loaded ones was found to be sufficient for providing comparable coverage and can be clinically used in eye-cancer-busy centers.
Identifiants
pubmed: 37738654
doi: 10.1002/acm2.14149
pmc: PMC10691642
doi:
Substances chimiques
Water
059QF0KO0R
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
e14149Informations de copyright
© 2023 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine.
Références
Clin Dermatol. 2009 Jan-Feb;27(1):122-33
pubmed: 19095158
Brachytherapy. 2008 Oct-Dec;7(4):327-35
pubmed: 18782684
Med Phys. 2012 Oct;39(10):6161-84
pubmed: 23039655
Radiat Oncol. 2018 Nov 13;13(1):221
pubmed: 30424782
Med Phys. 2014 Feb;41(2):021717
pubmed: 24506608
Med Phys. 1993 Mar-Apr;20(2 Pt 1):383-9
pubmed: 8497229
Brachytherapy. 2019 Sep - Oct;18(5):658-667
pubmed: 31235445
Med Phys. 2021 Aug;48(8):4542-4559
pubmed: 34250607
Med Phys. 2012 Jun;39(6):3332-42
pubmed: 22755715
Phys Med Biol. 2016 Dec 7;61(23):8214-8231
pubmed: 27804922
J Appl Clin Med Phys. 2023 Dec;24(12):e14149
pubmed: 37738654
Med Phys. 2010 Jan;37(1):368-76
pubmed: 20175499
Med Phys. 2011 Jan;38(1):306-16
pubmed: 21361199
J Appl Clin Med Phys. 2020 Sep;21(9):57-70
pubmed: 32656945
Med Phys. 2008 Jul;35(7):3364-71
pubmed: 18697560
Med Phys. 2008 Dec;35(12):5530-43
pubmed: 19175111
J Contemp Brachytherapy. 2012 Sep;4(3):165-75
pubmed: 23346146