Ultra-high performance supercritical fluid chromatography coupled to tandem mass spectrometry for antidoping analyses: Assessment of the inter-laboratory reproducibility with urine samples.
anti‐doping analyses
inter‐laboratory reproducibility
tandem mass spectrometry
ultra high‐performance supercritical fluid chromatography
Journal
Analytical science advances
ISSN: 2628-5452
Titre abrégé: Anal Sci Adv
Pays: Germany
ID NLM: 9918334986206676
Informations de publication
Date de publication:
Feb 2021
Feb 2021
Historique:
received:
26
10
2020
revised:
26
11
2020
accepted:
26
11
2020
medline:
5
12
2020
pubmed:
5
12
2020
entrez:
8
5
2024
Statut:
epublish
Résumé
The aim of this study was to assess the interlaboratory reproducibility of ultra-high performance supercritical fluid chromatography coupled with tandem mass spectrometry method for routine antidoping analyses. To do so, a set of 21 doping agents, spiked in urine and analyzed after dilute and shoot treatment, was used to assess the variability of their retention times between four different laboratories, all equipped with the same chromatographic system and with the same ultra-high performance supercritical fluid chromatography stationary phase chemistry. The average relative standard deviations (RSD%) demonstrated a good reproducibility of the retention times for 19 out of 21 analytes, with RSD% values below 3.0%. Only for two substances, namely fenbutrazate and niketamide, the retention was not repeatable between laboratories, with RSD% of approximately 15% in both cases. This behaviour was associated with (a) the low organic modifier percentage (around 2-4%) in the mobile phase at the corresponding retention times, and (b) the influence of the system volume on poorly retained analytes. An analysis on seven "blind" urines was subsequently carried out in the same four laboratories. In these blind samples, either one, two, or none of the 21 doping agents previously analyzed were present at an unknown concentration. Each laboratory had to perform the identification of the compounds in the samples and estimate their concentrations. All laboratories assigned all target analytes correctly in all blind urine samples and provide a comparable estimation of their concentrations.
Identifiants
pubmed: 38715742
doi: 10.1002/ansa.202000131
pii: ANSA202000131
pmc: PMC10989559
doi:
Types de publication
Journal Article
Langues
eng
Pagination
68-75Informations de copyright
© 2020 The Authors. Analytical Science Advances published by Wiley‐VCH GmbH.