[Repeatability of spirometrys expiratory flows].
Repetitividad de los flujos espiratorios de la espirometría.
Forced Expiratory Volume
Reproducibility of Results
Respiratory Function Tests
Respiratory Physiological Phenomena
Spirometry
Journal
Revista medica del Instituto Mexicano del Seguro Social
ISSN: 2448-5667
Titre abrégé: Rev Med Inst Mex Seguro Soc
Pays: Mexico
ID NLM: 101243727
Informations de publication
Date de publication:
08 Jan 2024
08 Jan 2024
Historique:
received:
20
06
2023
accepted:
20
09
2023
medline:
8
8
2024
pubmed:
7
8
2024
entrez:
7
8
2024
Statut:
epublish
Résumé
The quality of the spirometry is estimated with criteria of acceptability and repeatability. The repeatability criteria accepted by consensus is < 0.150 L. To know the repeatability in quality A spirometry. Analytical cross-sectional design. The demographic variables and the 3 best spirometry curves with normal, suggestive of restriction and bronchial obstruction profiles were obtained from consecutive subjects of both genders from 18 to 80 years of age. The repeatability was analyzed with the mean difference (bias) and the intraclass correlation coefficient. 630 curves from 210 subjects were accepted. Group age 60 ± 15 years. Female predominance 113 (53.8%), occupation: domestic services 61 (29%), and diagnosed with chronic obstructive pulmonary disease: 70 (33.4%). The differences in the curves were < 0.150 L. The mean difference (bias) and the intraclass correlation coefficient (95% confidence interval, 95% CI) of the forced expiratory volume in the first second were 1 vs. 2 maneuver: -0.01 (0.13, -0.14), 0.997 (95% CI 0.996, 0.998); 2 vs. 3 maneuver: 0.00 (0.13, -0.13), 0.997 (95% CI 0.996, 0.998), and maneuver 1 vs. 3: -0.00 (0.16, -0.17), 0.995 (95% CI 0.994, 0.996). Forced vital capacity: 1 vs. 2 maneuver: -0.01 (0.17, -0.18), 0.996 (95% CI 0.995, 0.997); 2 vs. 3 maneuver: 0.01 (0.17, -0.16), 0.997 (95% CI 0.0.996, 0.998), and maneuver 1 vs. 3: -0.00 (0.18, -0.19), 0.996 (95% CI 0.995, 0.997). The repeatability obtained in spirometry with quality A validates the use of the repeatability criterion of 0.150 L. la calidad de la espirometría se estima con criterios de aceptabilidad y repetitividad. La repetitividad aceptada por consenso es < 0.150 L. conocer la repetitividad en espirometrías de calidad A. diseño transversal analítico. Se obtuvieron las variables demográficas y las 3 mejores curvas de espirometría con perfil normal, que sugiriera restricción y obstrucción bronquial de sujetos consecutivos de ambos géneros de 18 a 80 años. La repetitividad se analizó con la diferencia de medias (sesgo) y el coeficiente de correlación intraclase. se aceptaron 630 curvas de 210 sujetos. Edad grupal 60 ± 15 años. Predominio femenino 113 (53.8%), ocupación: servicios domésticos 61 (29%) y con enfermedad pulmonar obstructiva crónica 70 (33.4%). Las diferencias en las curvas fueron < 0.150 L. Las diferencias medias (sesgo) y el coeficiente de correlación intraclase (intervalo de confianza al 95%, IC 95%) del volumen espiratorio forzado en el primer segundo fueron: maniobra 1 frente a 2: −0.01 (0.13, −0.14), 0.997 (IC 95% 0.996, 0.998); maniobra 2 frente a 3: 0.00 (0.13, −0.13), 0.997 (IC 95% 0.996, 0.998), y maniobra 1 frente a 3: −0.00 (0.16, −0.17), 0.995 (IC 95% 0.994, 0.996). La capacidad vital forzada: maniobra 1 frente a 2: −0.01 (0.17, −0.18), 0.996 (IC 95% 0.995, 0.997); maniobra 2 frente a 3: 0.01 (0.17, −0.16), 0.997 (IC 95% 0.0.996, 0.998), y maniobra 1 frente a 3: −0.00 (0.18, −0.19), 0.996 (IC 95% 0.995, 0.997). la repetitividad obtenida en espirometrías con calidad A valida el uso del criterio de repetitividad de 0.150 L.
Sections du résumé
Background
UNASSIGNED
The quality of the spirometry is estimated with criteria of acceptability and repeatability. The repeatability criteria accepted by consensus is < 0.150 L.
Objective
UNASSIGNED
To know the repeatability in quality A spirometry.
Material and methods
UNASSIGNED
Analytical cross-sectional design. The demographic variables and the 3 best spirometry curves with normal, suggestive of restriction and bronchial obstruction profiles were obtained from consecutive subjects of both genders from 18 to 80 years of age. The repeatability was analyzed with the mean difference (bias) and the intraclass correlation coefficient.
Results
UNASSIGNED
630 curves from 210 subjects were accepted. Group age 60 ± 15 years. Female predominance 113 (53.8%), occupation: domestic services 61 (29%), and diagnosed with chronic obstructive pulmonary disease: 70 (33.4%). The differences in the curves were < 0.150 L. The mean difference (bias) and the intraclass correlation coefficient (95% confidence interval, 95% CI) of the forced expiratory volume in the first second were 1 vs. 2 maneuver: -0.01 (0.13, -0.14), 0.997 (95% CI 0.996, 0.998); 2 vs. 3 maneuver: 0.00 (0.13, -0.13), 0.997 (95% CI 0.996, 0.998), and maneuver 1 vs. 3: -0.00 (0.16, -0.17), 0.995 (95% CI 0.994, 0.996). Forced vital capacity: 1 vs. 2 maneuver: -0.01 (0.17, -0.18), 0.996 (95% CI 0.995, 0.997); 2 vs. 3 maneuver: 0.01 (0.17, -0.16), 0.997 (95% CI 0.0.996, 0.998), and maneuver 1 vs. 3: -0.00 (0.18, -0.19), 0.996 (95% CI 0.995, 0.997).
Conclusion
UNASSIGNED
The repeatability obtained in spirometry with quality A validates the use of the repeatability criterion of 0.150 L.
Introducción
UNASSIGNED
la calidad de la espirometría se estima con criterios de aceptabilidad y repetitividad. La repetitividad aceptada por consenso es < 0.150 L.
Objetivo
UNASSIGNED
conocer la repetitividad en espirometrías de calidad A.
Material y métodos
UNASSIGNED
diseño transversal analítico. Se obtuvieron las variables demográficas y las 3 mejores curvas de espirometría con perfil normal, que sugiriera restricción y obstrucción bronquial de sujetos consecutivos de ambos géneros de 18 a 80 años. La repetitividad se analizó con la diferencia de medias (sesgo) y el coeficiente de correlación intraclase.
Resultados
UNASSIGNED
se aceptaron 630 curvas de 210 sujetos. Edad grupal 60 ± 15 años. Predominio femenino 113 (53.8%), ocupación: servicios domésticos 61 (29%) y con enfermedad pulmonar obstructiva crónica 70 (33.4%). Las diferencias en las curvas fueron < 0.150 L. Las diferencias medias (sesgo) y el coeficiente de correlación intraclase (intervalo de confianza al 95%, IC 95%) del volumen espiratorio forzado en el primer segundo fueron: maniobra 1 frente a 2: −0.01 (0.13, −0.14), 0.997 (IC 95% 0.996, 0.998); maniobra 2 frente a 3: 0.00 (0.13, −0.13), 0.997 (IC 95% 0.996, 0.998), y maniobra 1 frente a 3: −0.00 (0.16, −0.17), 0.995 (IC 95% 0.994, 0.996). La capacidad vital forzada: maniobra 1 frente a 2: −0.01 (0.17, −0.18), 0.996 (IC 95% 0.995, 0.997); maniobra 2 frente a 3: 0.01 (0.17, −0.16), 0.997 (IC 95% 0.0.996, 0.998), y maniobra 1 frente a 3: −0.00 (0.18, −0.19), 0.996 (IC 95% 0.995, 0.997).
Conclusión
UNASSIGNED
la repetitividad obtenida en espirometrías con calidad A valida el uso del criterio de repetitividad de 0.150 L.
Autres résumés
Type: Publisher
(spa)
la calidad de la espirometría se estima con criterios de aceptabilidad y repetitividad. La repetitividad aceptada por consenso es < 0.150 L.
Identifiants
pubmed: 39110813
doi: 10.5281/zenodo.10278119
doi:
pii:
Types de publication
English Abstract
Journal Article
Observational Study
Langues
spa
Sous-ensembles de citation
IM
Pagination
1-9Informations de copyright