The Japan Network for Research and Information on Medical Exposures (J-RIME) established the diagnostic reference level (DRL) and is advancing optimization of radiation protection. We believe that the...
To explore the potentials of adding copper (Cu) filter on image quality and patient dose of adult patients underwent chest X-ray examination METHODS: Patients were divided into four groups. Group 1, p...
EI, ESD and DAP data for a total of 784 patients (422 male and 362 female) that underwent indirect digital chest radiography exam were collected. Image quality was maintained when adding 0.1 mm Cu fil...
Reducing dose when using indirect digital chest radiography is possible with no trade-off on image quality. No loss of image quality was reported, images were broadly comparable....
This study highlights the importance of utilizing the additional copper filter in digital chest radiography....
We have been using a paper-based hard copy print (paper print) system of X-ray images, in which digital imaging and communications in medicine (DICOM) data can be directly output on papers from medica...
We measured contrast response to verify consistency of image appearance on both display systems. The contrast resolution was assessed by a CDRAD phantom. The spatial resolution was assessed by an X-ra...
The contrast response of the paper printer was not concordant with the grayscale standard display function (GSDF). The difference between the measured contrast response and the ideal GSDF on the paper...
The display performance of the paper printer for X-ray images was clarified. X-ray images printed on the paper should be used carefully taking account of their characteristics of display performance....
Computerized image analysis is a viable technique for evaluating image quality as a complement to human observers....
To systematically review the image analysis software used in the assessment of 2D image quality using mammography phantoms....
A systematic search of multiple databases was performed from inception to July 2020 for articles that incorporated computerized analysis of 2D images of physical mammography phantoms to determine imag...
A total of 26 studies were included, 12 were carried out using direct digital imaging and 14 using screen film mammography. The ACR phantom (model-156) was the most frequently evaluated phantom, possi...
Image analysis software was in agreement with human observers but demonstrated higher consistency and reproducibility of quality evaluation. Additionally, using computerized analysis, several quantita...
The proportion of diagnostic images not applied for diagnostic purposes is an indicator of image quality, safety, and efficiency in radiography. Despite increased awareness, image reject is still a su...
All skeletal images at two digital X-ray rooms at two public hospitals in Norway were reviewed for four weeks in 2020. The number of exposed images, type of examination, and number of deleted images w...
2183 and 1467 X-ray images were taken at Hospital 1 and 2 respectively. The corresponding reject rates were 14.2% and 9.1%. The reject rate varied greatly from day to day (from 0% to 22%), and the exa...
The reject rate is high and reduces quality, safety, and efficiency of imaging services. The reasons for image rejects are typical professionally reducible errors indicating great potential for improv...
Monitoring and assessing image rejects are of great importance to management, training, education, patient safety, and for quality improvement of imaging services....
To investigate the effect of the exposure parameters on image quality (IQ) metrics of phantom images, obtained automatically using software or from visual evaluation....
Three commercial phantoms and a homemade phantom constructed according to the instructions given in the IAEA Human Health Series No. 39 publication were used, along with the respective software that e...
The effects of the exposure parameters on the IQ metrics, determined with the commercial and the IAEA phantoms, were not the same. For example, clear trends of improvement of IQ scores with increased ...
Failure to detect some expected trends with the complex commercial phantoms may be attributed to the fact that IQ in DR is more strongly affected by the post-processing procedures, which may mask the ...
Pectoral muscle removal is a fundamental preliminary step in computer-aided diagnosis systems for full-field digital mammography (FFDM). Currently, two open-source publicly available packages (LIBRA a...
To compare performance of the two packages on a single database of FFDM images....
Only mediolateral oblique (MLO) FFDM was considered because of large presence of pectoral muscle on this type of projection. For obtaining ground truth, pectoral muscle has been manually segmented by ...
FFDMs from 168 consecutive women at our institution have been included in the study. Both LIBRA's Dice-index and Cohen-kappa were significantly higher than OpenBreast (Wilcoxon signed-rank test P < 0....
Libra has a better performance than OpenBreast in pectoral muscle delineation so that, although our study has not a direct clinical application, these results are useful in the choice of packages for ...
The local dense area (LDA) test is a test for evaluating the stability of image quality published in the EUREF fourth edition-supplements for local high-density areas. In this test, a small PMMA plate...
Repeated chest and/or abdomen radiographs are needed on mobile digital radiography (DR) units to assess infants' health status. Optimization of DR tube peak kilovoltage (kVp) and tube current time pro...
To investigate the effect of exposure parameters and additional filtration on entrance skin dose and image quality in newborn DR imaging....
A physical anthropomorphic phantom that simulates the average full-term neonate was used. Chest and chest/abdomen DR images were acquired at the manufacturer's recommended kVp/mAs exposure parameters ...
Signal difference increased with kVp and progressively decreased with increasing filtration. Compared to the manufacturer's recommended 53 kVp/1.6 mAs exposure parameters, ESD was reduced by 76% (47.6...
The results derived from this phantom study suggest that additional beam filtration along with the appropriate leverage of exposure parameters can lower ESD while maintaining image quality in full-ter...