Radiation dose reduction for musculoskeletal computed tomography of the pelvis with preserved image quality.


Journal

Skeletal radiology
ISSN: 1432-2161
Titre abrégé: Skeletal Radiol
Pays: Germany
ID NLM: 7701953

Informations de publication

Date de publication:
Mar 2019
Historique:
received: 12 12 2017
accepted: 30 07 2018
revised: 23 07 2018
pubmed: 30 8 2018
medline: 2 2 2019
entrez: 30 8 2018
Statut: ppublish

Résumé

To analyze the impact of pelvic computed tomography (CT) technique optimization on estimated dose and subjective and objective image quality. An institutional review board (IRB)-approved retrospective records review was performed with waived informed consent. Five CT scanners (various manufacturers/models) were standardized to match the lowest dose profile on campus via subjective assessment of clinical images by experienced musculoskeletal radiologists. The lowest dose profile had previously been established through image assessment by experienced musculoskeletal radiologists after a department-wide radiation dose reduction initiative. A consecutive series of 60 pre- and 59 post-optimization bony pelvis CTs were analyzed by two residents, who obtained signal-to-noise ratio for femoral cortex and marrow, gluteus medius muscle, and subcutaneous and visceral fat in a standardized fashion. Two blinded attending radiologists ranked image quality from poor to excellent. Pre- and post-optimization subjects exhibited no difference in gender, age, or BMI (p > 0.2). Mean CT dose index (CTDIvol) and dose-length product (DLP) decreased by approximately 45%, from 39± 14 to 18± 12 mGy (p < 0.0001) and 1,227± 469 to 546± 384 mGy-cm (p < 0.0001). Lower body mass index (BMI) was associated with a larger dose reduction and higher BMI with higher DLP regardless of pre- or post-optimization examination. Inter-observer agreement was 0.64-0.92 for SNR measurements. Cortex SNR increased significantly for both observers (p < 0.02). Although qualitative image quality significantly decreased for one observer (p < 0.01), adequate mean quality (3.3 out of 5) was maintained for both observers. Subjective and objective image quality for pelvic CT examination remains adequate, despite a substantially reduced radiation dose.

Identifiants

pubmed: 30155628
doi: 10.1007/s00256-018-3039-9
pii: 10.1007/s00256-018-3039-9
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

375-385

Références

Br J Radiol. 2004;77 Spec No 1:S46-63
pubmed: 15546842
Med Phys. 1992 Mar-Apr;19(2):325-7
pubmed: 1584125
Radiology. 2006 Sep;240(3):828-34
pubmed: 16837668
Radiology. 2011 May;259(2):565-73
pubmed: 21386048
Tutor Quant Methods Psychol. 2012;8(1):23-34
pubmed: 22833776
J Thorac Imaging. 2013 Jan;28(1):40-8
pubmed: 22847638
Radiology. 1990 Oct;177(1):83-8
pubmed: 2399343
Eur J Radiol. 2014 Oct;83(10):1953-61
pubmed: 25066756
Clin Radiol. 2015 Feb;70(2):176-82
pubmed: 25496824
J Trauma. 1989 Dec;29(12):1643-6
pubmed: 2593194
Health Phys. 2016 Mar;110(3):260-1
pubmed: 26808877
Eur Radiol. 2016 Oct;26(10):3608-16
pubmed: 26847044
Acad Radiol. 2016 Apr;23(4):438-45
pubmed: 26872868
Br J Radiol. 2016;89(1060):20150082
pubmed: 26882825
Cancer. 1968 Oct;22(4):767-78
pubmed: 5212301

Auteurs

Elaina Zabak (E)

UT Southwestern Medical Center, 5323 Harry Hines Blvd, Dallas, TX, 75390-9178, USA.

Hythem Omar (H)

UT Southwestern Medical Center, 5323 Harry Hines Blvd, Dallas, TX, 75390-9178, USA.

Ethan Boothe (E)

UT Southwestern Medical Center, 5323 Harry Hines Blvd, Dallas, TX, 75390-9178, USA. Ethan.Boothe@UTSouthwestern.edu.

Lulu Tenorio (L)

UT Southwestern Medical Center, 5323 Harry Hines Blvd, Dallas, TX, 75390-9178, USA.

Jeffrey Guild (J)

UT Southwestern Medical Center, 5323 Harry Hines Blvd, Dallas, TX, 75390-9178, USA.

Suhny Abbara (S)

UT Southwestern Medical Center, 5323 Harry Hines Blvd, Dallas, TX, 75390-9178, USA.

Avneesh Chhabra (A)

UT Southwestern Medical Center, 5323 Harry Hines Blvd, Dallas, TX, 75390-9178, USA.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH