Experimental and Monte-Carlo characterization of the novel compact ionization chamber PTW 31023 for reference and relative dosimetry in high energy photon beams.

Compact ionization chamber Correction factors Photon-beam dosimetry

Journal

Zeitschrift fur medizinische Physik
ISSN: 1876-4436
Titre abrégé: Z Med Phys
Pays: Germany
ID NLM: 100886455

Informations de publication

Date de publication:
Dec 2019
Historique:
received: 05 09 2018
revised: 20 02 2019
accepted: 20 02 2019
pubmed: 18 3 2019
medline: 6 5 2020
entrez: 18 3 2019
Statut: ppublish

Résumé

The aim of the present work is to perform dosimetric characterization of a novel vented PinPoint ionization chamber (PTW 31023, PTW-Freiburg, Germany). This chamber replaces the previous model (PTW 31014), where the diameter of the central electrode has been increased from 0.3 to 0.6mm and the guard ring has been redesigned. Correction factors for reference and non-reference measurement conditions were examined. Measurements and calculations of the correction factors were performed according to the DIN 6800-2. The shifts of the effective point of measurement (EPOM) from the chamber's reference point were determined by comparison of the measured PDD with the reference curve obtained with a Roos chamber. Its lateral dose response functions, which act according to a mathematical convolution model as the convolution kernel transforming the dose profile D(x) to the measured signal M(x), have been approximated by Gaussian functions with standard deviation σ. Additionally, the saturation correction factors k The shifts of the EPOM from the reference point, Δz, are found to be -0.55 (6MV) and -0.56 (10MV) in the radial orientation and -0.97mm (6MV) and -0.91mm (10MV) in the axial orientation. All values of Δz have an uncertainty of 0.1mm. The σ values are 0.80mm (axial), 0.75mm (radial lateral) and 1.76mm (radial longitudinal) for 6MV photon beam and are 0.85mm (axial), 0.75mm (radial lateral) and 1.82mm (radial longitudinal) for 15MV photon beam. All σ values have an uncertainty of 0.05mm. The correction factor k Correction factors associated with the new chamber required to perform reference and relative dose measurements have been determined according to the DIN-protocol. The correction factor k

Identifiants

pubmed: 30878324
pii: S0939-3889(18)30113-2
doi: 10.1016/j.zemedi.2019.02.002
pii:
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

303-313

Informations de copyright

Copyright © 2019. Published by Elsevier GmbH.

Auteurs

Isabel Büsing (I)

University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, Germany. Electronic address: isabel.buesing@uni-oldenburg.de.

Andre Brant (A)

University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, Germany.

Tobias Lange (T)

University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, Germany.

Björn Delfs (B)

University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, Germany.

Daniela Poppinga (D)

PTW-Freiburg, Freiburg, Germany.

Rafael Kranzer (R)

PTW-Freiburg, Freiburg, Germany.

Hui Khee Looe (HK)

University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, Germany.

Björn Poppe (B)

University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, Germany.

Articles similaires

Humans Female Precision Medicine Radiotherapy Planning, Computer-Assisted Breast Neoplasms
Monte Carlo Method Models, Neurological Neuronal Plasticity Computational Biology Humans
Humans Proton Therapy Lung Neoplasms Radiotherapy Planning, Computer-Assisted Radiotherapy, Intensity-Modulated
Monte Carlo Method Fiber Optic Technology Scattering, Radiation Spectrum Analysis Equipment Design

Classifications MeSH