Experimental and Monte-Carlo characterization of the novel compact ionization chamber PTW 31023 for reference and relative dosimetry in high energy photon beams.
Compact ionization chamber
Correction factors
Photon-beam dosimetry
Journal
Zeitschrift fur medizinische Physik
ISSN: 1876-4436
Titre abrégé: Z Med Phys
Pays: Germany
ID NLM: 100886455
Informations de publication
Date de publication:
Dec 2019
Dec 2019
Historique:
received:
05
09
2018
revised:
20
02
2019
accepted:
20
02
2019
pubmed:
18
3
2019
medline:
6
5
2020
entrez:
18
3
2019
Statut:
ppublish
Résumé
The aim of the present work is to perform dosimetric characterization of a novel vented PinPoint ionization chamber (PTW 31023, PTW-Freiburg, Germany). This chamber replaces the previous model (PTW 31014), where the diameter of the central electrode has been increased from 0.3 to 0.6mm and the guard ring has been redesigned. Correction factors for reference and non-reference measurement conditions were examined. Measurements and calculations of the correction factors were performed according to the DIN 6800-2. The shifts of the effective point of measurement (EPOM) from the chamber's reference point were determined by comparison of the measured PDD with the reference curve obtained with a Roos chamber. Its lateral dose response functions, which act according to a mathematical convolution model as the convolution kernel transforming the dose profile D(x) to the measured signal M(x), have been approximated by Gaussian functions with standard deviation σ. Additionally, the saturation correction factors k The shifts of the EPOM from the reference point, Δz, are found to be -0.55 (6MV) and -0.56 (10MV) in the radial orientation and -0.97mm (6MV) and -0.91mm (10MV) in the axial orientation. All values of Δz have an uncertainty of 0.1mm. The σ values are 0.80mm (axial), 0.75mm (radial lateral) and 1.76mm (radial longitudinal) for 6MV photon beam and are 0.85mm (axial), 0.75mm (radial lateral) and 1.82mm (radial longitudinal) for 15MV photon beam. All σ values have an uncertainty of 0.05mm. The correction factor k Correction factors associated with the new chamber required to perform reference and relative dose measurements have been determined according to the DIN-protocol. The correction factor k
Identifiants
pubmed: 30878324
pii: S0939-3889(18)30113-2
doi: 10.1016/j.zemedi.2019.02.002
pii:
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
303-313Informations de copyright
Copyright © 2019. Published by Elsevier GmbH.