Technical note: accuracy and precision in stereotactic stem cell transplantation.


Journal

Acta neurochirurgica
ISSN: 0942-0940
Titre abrégé: Acta Neurochir (Wien)
Pays: Austria
ID NLM: 0151000

Informations de publication

Date de publication:
10 2019
Historique:
received: 25 03 2019
accepted: 24 05 2019
pubmed: 6 7 2019
medline: 9 6 2020
entrez: 6 7 2019
Statut: ppublish

Résumé

While multiple trials have employed stereotactic stem cell transplantation, injection techniques have received little critical attention. Precise cell delivery is critical for certain applications, particularly when targeting deep nuclei. Ten patients with a history of ischemic stroke underwent CT-guided stem cell transplantation. Cells were delivered along 3 tracts adjacent to the infarcted area. Intraoperative air deposits and postoperative T2-weighted MRI fluid signals were mapped in relation to calculated targets. The deepest air deposit was found 4.5 ± 1.0 mm (mean ± 2 SEM) from target. The apex of the T2-hyperintense tract was found 2.8 ± 0.8 mm from target. On average, air pockets were found anterior (1.2 ± 1.1 mm, p = 0.04) and superior (2.4 ± 1.0 mm, p < 0.001) to the target; no directional bias was noted for the apex of the T2-hyperintense tract. Location and distribution of air deposits were variable and were affected by the relationship of cannula trajectory to stroke cavity. Precise stereotactic cell transplantation is a little-studied technical challenge. Reflux of cell suspension and air, and the structure of the injection tract affect delivery of cell suspensions. Intraoperative CT allows assessment of delivery and potential trajectory correction.

Sections du résumé

BACKGROUND
While multiple trials have employed stereotactic stem cell transplantation, injection techniques have received little critical attention. Precise cell delivery is critical for certain applications, particularly when targeting deep nuclei.
METHODS
Ten patients with a history of ischemic stroke underwent CT-guided stem cell transplantation. Cells were delivered along 3 tracts adjacent to the infarcted area. Intraoperative air deposits and postoperative T2-weighted MRI fluid signals were mapped in relation to calculated targets.
RESULTS
The deepest air deposit was found 4.5 ± 1.0 mm (mean ± 2 SEM) from target. The apex of the T2-hyperintense tract was found 2.8 ± 0.8 mm from target. On average, air pockets were found anterior (1.2 ± 1.1 mm, p = 0.04) and superior (2.4 ± 1.0 mm, p < 0.001) to the target; no directional bias was noted for the apex of the T2-hyperintense tract. Location and distribution of air deposits were variable and were affected by the relationship of cannula trajectory to stroke cavity.
CONCLUSIONS
Precise stereotactic cell transplantation is a little-studied technical challenge. Reflux of cell suspension and air, and the structure of the injection tract affect delivery of cell suspensions. Intraoperative CT allows assessment of delivery and potential trajectory correction.

Identifiants

pubmed: 31273445
doi: 10.1007/s00701-019-03964-8
pii: 10.1007/s00701-019-03964-8
doi:

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

2059-2064

Auteurs

David Satzer (D)

Department of Neurosurgery, University of Chicago, 5841 S. Maryland Avenue, MC 3026, Chicago, IL, 60637, USA. david.satzer@uchospitals.edu.

Peter C Warnke (PC)

Department of Neurosurgery, University of Chicago, 5841 S. Maryland Avenue, MC 3026, Chicago, IL, 60637, USA.

Articles similaires

[Redispensing of expensive oral anticancer medicines: a practical application].

Lisanne N van Merendonk, Kübra Akgöl, Bastiaan Nuijen
1.00
Humans Antineoplastic Agents Administration, Oral Drug Costs Counterfeit Drugs

Smoking Cessation and Incident Cardiovascular Disease.

Jun Hwan Cho, Seung Yong Shin, Hoseob Kim et al.
1.00
Humans Male Smoking Cessation Cardiovascular Diseases Female
Humans United States Aged Cross-Sectional Studies Medicare Part C
1.00
Humans Yoga Low Back Pain Female Male

Classifications MeSH