Direct loading of blood for plasma separation and diagnostic assays on a digital microfluidic device.
Journal
Lab on a chip
ISSN: 1473-0189
Titre abrégé: Lab Chip
Pays: England
ID NLM: 101128948
Informations de publication
Date de publication:
19 05 2020
19 05 2020
Historique:
pubmed:
28
4
2020
medline:
22
6
2021
entrez:
28
4
2020
Statut:
ppublish
Résumé
Finger-stick blood sampling is convenient for point of care diagnostics, but whole blood samples are problematic for many assays because of severe matrix effects associated with blood cells and cell debris. We introduce a new digital microfluidic (DMF) diagnostic platform with integrated porous membranes for blood-plasma separation from finger-stick blood volumes, capable of performing complex, multi-step, diagnostic assays. Importantly, the samples can be directly loaded onto the device by a finger "dab" for user-friendly operation. We characterize the platform by comparison to plasma generated via the "gold standard" centrifugation technique, and demonstrate a 21-step rubella virus (RV) IgG immunoassay yielding a detection limit of 1.9 IU mL-1, below the diagnostic cut-off. We propose that this work represents a critical next step in DMF based portable diagnostic assays-allowing the analysis of whole blood samples without pre-processing.
Substances chimiques
Immunoglobulin G
0
Types de publication
Journal Article
Research Support, Non-U.S. Gov't
Langues
eng
Sous-ensembles de citation
IM
Pagination
1845-1855Subventions
Organisme : CIHR
Pays : Canada