Long-Term Outcomes of the Mosaic Aortic Porcine Bioprosthesis in Japan - Results From the Japan Mosaic Valve Long-Term Multicenter Study.
Aged
Aged, 80 and over
Animals
Aortic Valve
/ diagnostic imaging
Bioprosthesis
Female
Heart Valve Diseases
/ diagnostic imaging
Heart Valve Prosthesis
Heart Valve Prosthesis Implantation
/ adverse effects
Hemodynamics
Heterografts
Humans
Japan
Male
Middle Aged
Postoperative Complications
/ mortality
Prosthesis Failure
Recovery of Function
Retrospective Studies
Sus scrofa
Time Factors
Treatment Outcome
Aortic valve replacement
Bioprosthesis
Durability
Structural valve deterioration
Journal
Circulation journal : official journal of the Japanese Circulation Society
ISSN: 1347-4820
Titre abrégé: Circ J
Pays: Japan
ID NLM: 101137683
Informations de publication
Date de publication:
22 07 2020
22 07 2020
Historique:
pubmed:
20
6
2020
medline:
15
12
2020
entrez:
20
6
2020
Statut:
ppublish
Résumé
This study retrospectively evaluated the long-term patient outcomes and durability of the Mosaic aortic porcine bioprosthesis in the Japan Mosaic valve long-term multicenter study.Methods and Results:We reviewed the records of 1,202 patients who underwent aortic valve replacement with the Mosaic bioprosthesis at 10 centers in Japan (1999-2014). Patient data were collected using Research Electronic Data Capture. Patient survival was determined by Kaplan-Meier methodology. Freedom from structural valve deterioration (SVD) and valve-related reoperation and death were determined by actuarial methods. The median (interquartile range [IQR]) age of the cohort was 76 (70-80) years. The median (IQR) follow-up period was 3.52 (1.71-5.35) years. The longest follow-up was 15.8 years. The 30-day mortality rate was 2.3%. The 12-year actuarial survival rate was 59.9±7.5%, and the freedom from valve-related death was 81.1±7.9%. The freedom from reoperation was 86.4±2.6% at 12 years. The freedom from SVD at 12 years was 93.5±2.9% for patients aged ≥65 years and 98.2±1.8% for those aged <65 years. The median (IQR) systolic pressure gradient was 17 (12-23) and 19 (12-25) mmHg at 1 and 10 years, respectively. The median (IQR) effective orifice area was 1.2 (1.1-1.5) and 1.1 (1-1.5) cm The Mosaic porcine bioprosthesis showed satisfactory long-term outcomes over 12 years.
Sections du résumé
BACKGROUND
This study retrospectively evaluated the long-term patient outcomes and durability of the Mosaic aortic porcine bioprosthesis in the Japan Mosaic valve long-term multicenter study.Methods and Results:We reviewed the records of 1,202 patients who underwent aortic valve replacement with the Mosaic bioprosthesis at 10 centers in Japan (1999-2014). Patient data were collected using Research Electronic Data Capture. Patient survival was determined by Kaplan-Meier methodology. Freedom from structural valve deterioration (SVD) and valve-related reoperation and death were determined by actuarial methods. The median (interquartile range [IQR]) age of the cohort was 76 (70-80) years. The median (IQR) follow-up period was 3.52 (1.71-5.35) years. The longest follow-up was 15.8 years. The 30-day mortality rate was 2.3%. The 12-year actuarial survival rate was 59.9±7.5%, and the freedom from valve-related death was 81.1±7.9%. The freedom from reoperation was 86.4±2.6% at 12 years. The freedom from SVD at 12 years was 93.5±2.9% for patients aged ≥65 years and 98.2±1.8% for those aged <65 years. The median (IQR) systolic pressure gradient was 17 (12-23) and 19 (12-25) mmHg at 1 and 10 years, respectively. The median (IQR) effective orifice area was 1.2 (1.1-1.5) and 1.1 (1-1.5) cm
CONCLUSIONS
The Mosaic porcine bioprosthesis showed satisfactory long-term outcomes over 12 years.
Identifiants
pubmed: 32554951
doi: 10.1253/circj.CJ-19-1113
doi:
Types de publication
Journal Article
Multicenter Study
Observational Study
Langues
eng
Sous-ensembles de citation
IM