Investigating complement mediated interference in class I HLA-specific antibodies following renal transplantation.
Bacterial Proteins
/ metabolism
Biotin
/ analogs & derivatives
Cohort Studies
Complement System Proteins
/ metabolism
Edetic Acid
/ metabolism
Epitopes
/ immunology
Graft Rejection
/ diagnosis
HLA Antigens
/ immunology
Histocompatibility Antigens Class I
/ immunology
Histocompatibility Testing
/ methods
Humans
Isoantibodies
/ blood
Kidney Transplantation
Predictive Value of Tests
Prognosis
Retrospective Studies
Biotin-streptavidin
EDTA
Prozone
Renal transplant
SAB
Journal
Transplant immunology
ISSN: 1878-5492
Titre abrégé: Transpl Immunol
Pays: Netherlands
ID NLM: 9309923
Informations de publication
Date de publication:
10 2020
10 2020
Historique:
received:
17
05
2020
revised:
11
06
2020
accepted:
12
06
2020
pubmed:
24
6
2020
medline:
2
4
2021
entrez:
24
6
2020
Statut:
ppublish
Résumé
Single antigen bead testing (SAB) for HLA-specific antibody enables efficient organ allocation and aids in the diagnosis of antibody mediated rejection. In this retrospective cohort study, a population of kidney transplant recipients possessing HLA Class I antibodies was used to evaluate the best method for resolving complement interference, the so called "prozone" effect. The aim was to compare the use of EDTA versus a Biotin-Streptavidin Complex as methodological approaches for abating the prozone effect using a fixed 1 in 10 dilution as validation. One hundred and seventeen patients transplanted in our centre between 2009 and 2014 were identified as having class I HLA-specific antibody(-ies) using a Labscreen® Mixed assay. Positive sera underwent class I HLA-specific SAB testing; for comparison a standard SAB with and without EDTA, BSC and dilution (1 in 10) modifications were utilised. Samples were processed on the Luminex platform generating 11,349 bead reactions for analysis. We identified sera from 23 patients giving rise to 170 bead reactions showing complement interference. Using linear modelling, we observed slightly higher MFIs on average in both EDTA and BSC modifications when compared to the standard assay, allowing the nominal threshold MFI of 2000 in the standard assay to be adjusted to 2097 and 2033 in the EDTA and BSC assays respectively. We calculated 99% prediction intervals to establish outlier bead reactions for each assay. The 1 in 10 dilution was used as a crosscheck for determining which prozone reactions were overcome by EDTA and BSC. Using ROC curve analysis, EDTA was found to be ~90% sensitive and 100% specific compared to BSC which was ~60% sensitive and 100% specific in ameliorating prozone positive reactions at the thresholds defined by linear models. Our data indicates that both EDTA and BSC are suitable assays in overcoming CMI. We recommend that all clinical laboratories adopt a validated assay designed specifically to abrogate CMI for all potential renal transplant recipients, as the standard assay is inhibited in nearly 20% of a post-transplant cohort.
Identifiants
pubmed: 32574596
pii: S0966-3274(20)30085-X
doi: 10.1016/j.trim.2020.101310
pii:
doi:
Substances chimiques
Bacterial Proteins
0
Epitopes
0
HLA Antigens
0
Histocompatibility Antigens Class I
0
Isoantibodies
0
biotin-streptavidin complex
0
Biotin
6SO6U10H04
Complement System Proteins
9007-36-7
Edetic Acid
9G34HU7RV0
Types de publication
Comparative Study
Journal Article
Research Support, Non-U.S. Gov't
Langues
eng
Sous-ensembles de citation
IM
Pagination
101310Informations de copyright
Copyright © 2020 Elsevier B.V. All rights reserved.