Analytical investigation of ternary mixture of phenylephrine hydrochloride, dimetindene maleate and benzalkonium chloride using validated stability indicating HPLC-DAD method.
HPLC-DAD
Phenylephrine hydrochloride
benzalkonium chloride
dimetindene maleate
forced degradation
stability-indicating
topical nasal preparations
Journal
Drug development and industrial pharmacy
ISSN: 1520-5762
Titre abrégé: Drug Dev Ind Pharm
Pays: England
ID NLM: 7802620
Informations de publication
Date de publication:
Aug 2020
Aug 2020
Historique:
pubmed:
1
7
2020
medline:
12
3
2021
entrez:
30
6
2020
Statut:
ppublish
Résumé
A stability-indicating high performance liquid chromatography method with diode array detection (HPLC-DAD) was developed and validated for simultaneous determination of phenylephrine hydrochloride (PHR), dimetindene maleate (DMD) and benzalkonium chloride (BZM) in nasal drops and gel dosage forms. Effective liquid chromatographic separation was accomplished by employing Venusil XBP Cyano column (4.6 × 250 mm, 5 µm particle size) with gradient elution of the mobile phase consisting of buffer solution of potassium dihydrogen phosphate (0.025 M) and sodium 1-butane sulfonate (SBS) (0.025 M) (adjusted to pH 6.0) and acetonitrile. Peak areas of PHR, DMD and BZM at 271, 256 and 206 nm, respectively were measured and correlated to their concentrations. Peaks of PHR and DMD eluted at retention times 3.76 and 9.06 min, respectively, while BZM eluted as a couple of peaks at 11.88 and 12.51 min. The proposed HPLC procedure was carefully validated in terms of system suitability, linearity, ranges, precision, accuracy, specificity, robustness, detection and quantification limits. The linearity range for both PHR and BZM was 10-400 μg/mL and DMD was 5-300 μg/mL with correlation coefficients >0.9999. The studied compounds were subjected to stress conditions of neutral, acidic and alkaline hydrolysis, oxidation and thermal degradation. Good resolution of the three compounds from their forced degradation products proves specificity and stability-indicating merits of the proposed method. In addition, resolution of the three drugs under investigation from some pharmaceutical compounds of different medicinal categories showed the high specificity of the described method.
Identifiants
pubmed: 32594780
doi: 10.1080/03639045.2020.1788064
doi:
Substances chimiques
Benzalkonium Compounds
0
Phenylephrine
1WS297W6MV
Dimethindene
661FH77Z3P
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM