Limb Position Change Affects Tourniquet Pressure.
Journal
Journal of special operations medicine : a peer reviewed journal for SOF medical professionals
ISSN: 1553-9768
Titre abrégé: J Spec Oper Med
Pays: United States
ID NLM: 101158402
Informations de publication
Date de publication:
2021
2021
Historique:
accepted:
01
03
2021
entrez:
15
3
2021
pubmed:
16
3
2021
medline:
18
3
2021
Statut:
ppublish
Résumé
Limb position changes are likely during transport from injury location to definitive care. This study investigated passive limb position change effects on tourniquet pressure and occlusion. Triplicate buddy-applied OMNA® Marine Tourniquet applications to Doppler-based occlusion were done to sitting and laying supine mid-thigh (n=5) and sitting mid-arm (n=3). Tourniqueted limb positions were bent/straight/bent and straight/bent/straight (randomized first position order, 5 seconds/position, pressure every 0.1 second, two-way repeated measures ANOVA). Sitting thigh occlusion pressures leg bent were higher than straight (median, minimum-maximum; 328, 307-403mmHg versus 312, 295-387mmHg, p = .013). In each recipient, the pressure change for each position change for each limb had p < .003. In each recipient, when sitting, leg bent to straight increased pressure (326, 276-415mmHg to 371, 308-427mmHg bent first and 275, 233-354mmHg to 311, 241-353mmHg straight first), and straight to bent decreased pressure (371, 308-427mmHg to 301, 262-388mmHg bent first and 312, 265-395mmHg to 275, 233-354mmHg straight first). When laying, position changes from leg bent first resulted in pressure changes in each recipient but not in the same directions in each recipient. From laying leg straight first, in each recipient changing to bent increased the pressure (295, 210-366mmHg to 328, 255-376mmHg) and to straight decreased the pressure (328, 255-376 mmHg to 259, 210-333 mmHg). Sitting arm bent occlusion pressures were lower than straight (230, 228-252mmHg versus 256, 250-287mmHg, p = .026). Arm position changes resulted in pressure changes in each recipient but not in the same directions in each recipient. Changes in pressure trace character (presence or absence of rhythmically pulsatile traces) and Doppler-based occlusion were consistent with limb position-induced changes in tourniquet pressure (each p ≤ .001 leg, p = .071 arm traces, and p = .188 arm occlusion). Passive limb position changes can cause significant changes in tourniquet pressure. Therefore, tourniquet adequacy should be reassessed after any limb position change.
Sections du résumé
BACKGROUND
Limb position changes are likely during transport from injury location to definitive care. This study investigated passive limb position change effects on tourniquet pressure and occlusion.
METHODS
Triplicate buddy-applied OMNA® Marine Tourniquet applications to Doppler-based occlusion were done to sitting and laying supine mid-thigh (n=5) and sitting mid-arm (n=3). Tourniqueted limb positions were bent/straight/bent and straight/bent/straight (randomized first position order, 5 seconds/position, pressure every 0.1 second, two-way repeated measures ANOVA).
RESULTS
Sitting thigh occlusion pressures leg bent were higher than straight (median, minimum-maximum; 328, 307-403mmHg versus 312, 295-387mmHg, p = .013). In each recipient, the pressure change for each position change for each limb had p < .003. In each recipient, when sitting, leg bent to straight increased pressure (326, 276-415mmHg to 371, 308-427mmHg bent first and 275, 233-354mmHg to 311, 241-353mmHg straight first), and straight to bent decreased pressure (371, 308-427mmHg to 301, 262-388mmHg bent first and 312, 265-395mmHg to 275, 233-354mmHg straight first). When laying, position changes from leg bent first resulted in pressure changes in each recipient but not in the same directions in each recipient. From laying leg straight first, in each recipient changing to bent increased the pressure (295, 210-366mmHg to 328, 255-376mmHg) and to straight decreased the pressure (328, 255-376 mmHg to 259, 210-333 mmHg). Sitting arm bent occlusion pressures were lower than straight (230, 228-252mmHg versus 256, 250-287mmHg, p = .026). Arm position changes resulted in pressure changes in each recipient but not in the same directions in each recipient. Changes in pressure trace character (presence or absence of rhythmically pulsatile traces) and Doppler-based occlusion were consistent with limb position-induced changes in tourniquet pressure (each p ≤ .001 leg, p = .071 arm traces, and p = .188 arm occlusion).
CONCLUSIONS
Passive limb position changes can cause significant changes in tourniquet pressure. Therefore, tourniquet adequacy should be reassessed after any limb position change.
Identifiants
pubmed: 33721300
doi: 10.55460/H5BB-27ZC
doi:
pii:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
11-17Informations de copyright