Dosimetric accuracy of dual isocenter irradiation in low magnetic field resonance guided radiotherapy system for extended abdominal tumours.
Adaptive radiotherapy
Dual-isocenter
Extended-field IMRT
MRgRT
Journal
Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)
ISSN: 1724-191X
Titre abrégé: Phys Med
Pays: Italy
ID NLM: 9302888
Informations de publication
Date de publication:
Apr 2021
Apr 2021
Historique:
received:
23
11
2020
revised:
23
03
2021
accepted:
30
03
2021
pubmed:
26
4
2021
medline:
29
6
2021
entrez:
25
4
2021
Statut:
ppublish
Résumé
Due to limited field size of Magnetic Resonance Linear Accelerators (MR-Linac), some treatments could require a dual-isocenter planning approach to achieve a complete target coverage and thus exploit the benefits of the online adaptation. This study evaluates the dosimetric accuracy of the dual-isocenter intensity modulated radiation therapy (IMRT) delivery technique for MR-Linac. Dual-isocenter multi leaf collimator (MLC) and couch accuracy tests have been performed to evaluate the delivery accuracy of the system. A mono-isocenter plan delivered in clinical practice has then been retrospectively re-planned with dual-isocenter technique. The dual-isocenter plan has been re-calculated and delivered on a 3-dimensional (3D) ArcCHECK phantom and 2-dimensional (2D) films to assess its dosimetric accuracy in terms of gamma analysis. Clinical and planning target volume (CTV and PTV respectively) coverage robustness was then investigated after the introduction of ± 2 mm and ± 5 mm positioning errors by shifting the couch. MLC and couch accuracy tests confirmed the system accuracy in delivering a dual-isocenter irradiation. 2D/3D gamma analysis results occurred always to be above 95% if considered a gamma criteria 1%/2 mm and 1%/1 mm respectively for the 2D and 3D analysis. The mean variations for CTV D98% and PTV V95% were 0.2% and 1.1% respectively when positioning error was introduced separately in each direction, while the maximum observed variations were 0.9% (CTV) and 3.7% (PTV). The dosimetric accuracy of dual-isocenter irradiation has been verified for MR-Linac, achieving accurate and robust treatment strategy and improving dose conformality also in presence of targets whose extension exceeds the nominal maximum field size.
Identifiants
pubmed: 33895666
pii: S1120-1797(21)00155-1
doi: 10.1016/j.ejmp.2021.03.037
pii:
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
149-158Informations de copyright
Copyright © 2021 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
Déclaration de conflit d'intérêts
Declaration of Competing Interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.