Validation of a Custom Interface Pressure Measurement System to Improve Fitting of Transtibial Prosthetic Check Sockets.
amputation
interface pressure
prosthetic socket
Journal
Sensors (Basel, Switzerland)
ISSN: 1424-8220
Titre abrégé: Sensors (Basel)
Pays: Switzerland
ID NLM: 101204366
Informations de publication
Date de publication:
06 Apr 2023
06 Apr 2023
Historique:
received:
16
03
2023
revised:
31
03
2023
accepted:
04
04
2023
medline:
14
4
2023
entrez:
13
4
2023
pubmed:
14
4
2023
Statut:
epublish
Résumé
Achievement of fit between the residual limb and prosthetic socket during socket manufacture is a priority for clinicians and is essential for safety. Clinicians have recognised the potential benefits of having a sensor system that can provide objective socket-limb interface pressure measurements during socket fitting, but the cost of existing systems makes current technology prohibitive. This study will report on the characterisation, validation and preliminary clinical implementation of a low cost, portable, wireless sensor system designed for use during socket manufacture. Characterisation and benchtop testing demonstrated acceptable accuracy, behaviour at variable temperature, and dynamic response for use in prosthetic socket applications. Our sensor system was validated with simultaneous measurement by a commercial sensor system in the sockets of three transtibial prosthesis users during a fitting session in the clinic. There were no statistically significant differences between the sensor system and the commercial sensor for a variety of functional activities. The sensor system was found to be valid in this clinical context. Future work should explore how pressure data relates to ratings of fit and comfort, and how objective pressure data might be used to assist in clinical decision making.
Identifiants
pubmed: 37050838
pii: s23073778
doi: 10.3390/s23073778
pmc: PMC10099032
pii:
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Subventions
Organisme : University of Wollongong
ID : IV003
Références
Comput Methods Biomech Biomed Engin. 2012;15(3):295-301
pubmed: 21756121
Sensors (Basel). 2020 Nov 30;20(23):
pubmed: 33266213
Arch Rehabil Res Clin Transl. 2020 May 21;2(3):100059
pubmed: 33543086
Prosthet Orthot Int. 2020 Dec;44(6):384-401
pubmed: 33164655
Prosthet Orthot Int. 2023 Apr 1;47(2):194-203
pubmed: 36112475
Sci Transl Med. 2020 Dec 16;12(574):
pubmed: 33328330
Artif Limbs. 1962 Jun;6:4-15
pubmed: 13936729
Med Eng Phys. 2021 Nov;97:56-69
pubmed: 34756339
Acta Derm Venereol. 2011 Mar;91(2):173-7
pubmed: 21290085
Arch Phys Med Rehabil. 2017 Jan;98(1):105-113
pubmed: 27742450
Arch Phys Med Rehabil. 2011 Oct;92(10):1570-5
pubmed: 21963124
Front Rehabil Sci. 2022 Mar 07;3:853414
pubmed: 36189046
Arch Phys Med Rehabil. 2005 Apr;86(4):641-9
pubmed: 15827912
Prosthet Orthot Int. 2020 Feb;44(1):18-26
pubmed: 31769736
Prosthet Orthot Int. 2002 Dec;26(3):206-12
pubmed: 12562067
IEEE Trans Biomed Eng. 2018 Sep;65(9):1996-2010
pubmed: 29993506
Med Biol Eng Comput. 2018 Jul;56(7):1227-1240
pubmed: 29235055
J Rehabil Res Dev. 2008;45(1):15-29
pubmed: 18566923
J Rehabil Res Dev. 1999 Jul;36(3):155-63
pubmed: 10659798
Front Med Technol. 2022 Jun 17;4:908002
pubmed: 35782578