A Critique of Resonance Frequency Analysis and a Novel Method for Quantifying Dental Implant Stability in Vitro.


Journal

The International journal of oral & maxillofacial implants
ISSN: 1942-4434
Titre abrégé: Int J Oral Maxillofac Implants
Pays: United States
ID NLM: 8611905

Informations de publication

Date de publication:
Historique:
entrez: 27 2 2019
pubmed: 27 2 2019
medline: 18 12 2019
Statut: ppublish

Résumé

This study assessed the ability of resonance frequency measurements to differentiate the stability of implants with different lengths and diameters, and in different densities of bone. Another objective was to identify an alternative parameter capable of quantifying dental implant stability, thus facilitating greater sensitivity for efficacious detection of compromised or failing implants. Implants of two different diameters (4 and 5 mm) and six different lengths were individually placed in synthetic bone blocks of three different densities (15, 40/20, and 40 pounds per cubic foot) in combination with two different abutments (short and tall) to evaluate their stability. Resonance frequency measurements were obtained via Osstell ISQ and experimental modal analysis (EMA). The resonance frequency measurements were further confirmed via finite element analysis (FEA) using commercially available software ANSYS. Resonance frequencies measured via Osstell ISQ and EMA did not change with respect to the length of the implants. The FEA also confirmed the measured results. FEA simulations further indicated that angular stiffness at the neck of the implant (ie, the base of the abutment) varied considerably with respect to the implant length and diameter. Moreover, the calculated angular stiffness was independent of the type of abutment used. The results obtained from resonance frequency analyses did not accurately represent dental implant stability. Changes to implant length and diameter did not affect resonance frequencies. In contrast, angular stiffness at the neck of the implant represented a superior index for quantifying dental implant stability. It not only successfully differentiated stability of implants of both varying lengths and diameters, but also produced quantitative data that was independent of the type of abutments used.

Identifiants

pubmed: 30807625
doi: 10.11607/jomi.7173
doi:

Substances chimiques

Dental Implants 0

Types de publication

Journal Article

Langues

eng

Sous-ensembles de citation

IM

Pagination

595–603

Auteurs

Articles similaires

Humans Smartphone Blood Pressure Determination Oscillometry Male

Effect of soft tissue thickness on accuracy of conventional and digital implant impression techniques.

Eman Mostafa Awad, Mohamed Maamoun ElSheikh, Azza Abd El Moneim El-Segai
1.00
Dental Impression Technique Humans Dental Implants Computer-Aided Design Dental Impression Materials
Humans Male Middle Aged Female Adult
Pressure Water Computer Simulation Models, Theoretical Vibration

Classifications MeSH