Characterization of the Plastic Scintillator Detector System Exradin W2 in a High Dose Rate Flattening-Filter-Free Photon Beam.
Exradin W2
dosimetry
flattening-filter-free
high dose per pulse
high dose rate
output constancy
plastic scintillator
Journal
Sensors (Basel, Switzerland)
ISSN: 1424-8220
Titre abrégé: Sensors (Basel)
Pays: Switzerland
ID NLM: 101204366
Informations de publication
Date de publication:
08 Sep 2022
08 Sep 2022
Historique:
received:
28
07
2022
revised:
02
09
2022
accepted:
04
09
2022
entrez:
23
9
2022
pubmed:
24
9
2022
medline:
28
9
2022
Statut:
epublish
Résumé
(1) Background: The Exradin W2 is a commercially available scintillator detector designed for reference and relative dosimetry in small fields. In this work, we investigated the performance of the W2 scintillator in a 10 MV flattening-filter-free photon beam and compared it to the performance of ion chambers designed for small field measurements. (2) Methods: We measured beam profiles and percent depth dose curves with each detector and investigated the linearity of each system based on dose per pulse (DPP) and pulse repetition frequency. (3) Results: We found excellent agreement between the W2 scintillator and the ion chambers for beam profiles and percent depth dose curves. Our results also showed that the two-voltage method of calculating the ion recombination correction factor was sufficient to correct for the ion recombination effect of ion chambers, even at the highest DPP. (4) Conclusions: These findings show that the W2 scintillator shows excellent agreement with ion chambers in high DPP conditions.
Identifiants
pubmed: 36146135
pii: s22186785
doi: 10.3390/s22186785
pmc: PMC9505273
pii:
doi:
Substances chimiques
Plastics
0
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Subventions
Organisme : NCI NIH HHS
ID : P30 CA016672
Pays : United States
Références
J Appl Clin Med Phys. 2012 Nov 08;13(6):3803
pubmed: 23149774
J Appl Clin Med Phys. 2022 May;23(5):e13562
pubmed: 35157356
Phys Med Biol. 2015 Sep 7;60(17):6669-83
pubmed: 26271097
J Appl Clin Med Phys. 2015 May 08;16(3):5219
pubmed: 26103482
Med Phys. 2020 Mar;47(3):1327-1339
pubmed: 31860128
J Appl Clin Med Phys. 2022 Aug;23(8):e13728
pubmed: 35861648
Med Phys. 2019 May;46(5):2468-2476
pubmed: 30897221
Phys Med Biol. 1992 Oct;37(10):1901-13
pubmed: 1438555
Phys Med Biol. 2012 Nov 7;57(21):7133-45
pubmed: 23060036
Med Phys. 2014 Apr;41(4):041501
pubmed: 24694120
Phys Med Biol. 1992 Oct;37(10):1883-900
pubmed: 1438554
Med Phys. 2017 Feb;44(2):618-627
pubmed: 28001291
Med Phys. 2011 Apr;38(4):2140-50
pubmed: 21626947
Med Phys. 2006 Jan;33(1):128-35
pubmed: 16485419
Radiat Prot Dosimetry. 2006;120(1-4):1-6
pubmed: 16882685
Med Phys. 1999 Sep;26(9):1847-70
pubmed: 10505874
Phys Med Biol. 2000 Jul;45(7):1885-95
pubmed: 10943926
Phys Med Biol. 2012 May 7;57(9):2819-27
pubmed: 22510780